SAIT‐IFU‐GEMS‐H Rev.1.1 (2019‐11)
Page 66 of 79
Issue
Recommended action(s)
•
Restart the GEMS‐H Device. (Use the power switch located at the bottom
of the control pack).
•
If the above procedure does not solve the problem, the actuator may not
be in a normal state. Please contact the manufacturer of the device.
The following malfunctions:
‐
The assistive force suddenly
ceases without any reasons.
‐
The assistive forces are being
delivered to the patient, but hip
angles and torques are not
displayed on the connected
tablet PC.
‐
The device turns off
unexpectedly.
•
First, turn off and then take off the device.
•
Restart the GEMS‐H Device. (Use the power switch located at the bottom
of the control pack).
•
If the problem persists or occurs repeatedly, please contact the
manufacturer.
9.2. Device messages
Messages
Causes
Recommended action(s)
[Error] DETECTED ABNORMAL ACTUATOR
The physical therapist should attempt to
fix the problem by rebooting both the
device and the tablet PC. If this corrects
the error, the tablet PC will display
“
NORMAL OPERATION
”.
If rebooting the device and the tablet PC
do not clear the problems, please contact
the manufacturer.
[Error] DETECTED ABNORMAL IMU*
*IMU is used for the patient’s posture
estimation
Содержание GEMS-H
Страница 1: ...GEMS H Instructions for Use ...
Страница 19: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 19 of 79 4 6 Marking plate ...
Страница 38: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 38 of 79 Uncheck Unit US in order to use SI Units instead of US Units ...
Страница 52: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 52 of 79 6 7 3 3 Example of parameter control ...
Страница 54: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 54 of 79 Error message examples ...
Страница 64: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 64 of 79 ...
Страница 67: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 67 of 79 ...