SAIT‐IFU‐GEMS‐H Rev.1.1 (2019‐11)
Page 50 of 79
Parameter
GUI element
Function
GainL +/‐
Adjust assist strength of left leg only.
GainR +/‐
Adjust assist strength of right leg only.
Disable
Joint Limit
Turn off joint limit monitor (extension 40 degree
~
flexion 100
degree) which is used for safety functions.
Free
Motion
The device gives no assistive force, and moves itself just enough to
follow the patient as the patient moves.
Lock
Buttons
Disable all buttons to prevent accidental button presses.
Reset
Params
Reset all assist parameters back to initial values of the START point.
6.7.3.2. Parameter control monitoring
The therapist can monitor the control parameters and assist torque.