6.8.5. Mitigation of risks due to abnormal actuator module status
If the actuator module does not work properly during use, the controller may generate abnormal power. The
actuator module is monitored continuously during treatment, and when an abnormality is detected, the assistive
force is turned off.
Status of the device and recommended user action:
Should abnormal actuator status be detected, the tablet PC will display the error “
DETECTED ABNORMAL ACTUATOR
”
The actuator status is considered abnormal when data transmitted from the actuator does not match the expected
signals, for example, when the device detects patient movement, but actuator readings are constant. This may result
from disconnected communication lines, actuator faults, and/or software failure. The physical therapist should
attempt to fix the problem by rebooting both the device and the tablet PC. If this corrects the error, the tablet PC will
display “
NORMAL OPERATION
”.
If rebooting the device and the tablet PC do not clear the problems, please contact the manufacturer.
Error message examples
[Warning message]
[Screen of normal operation]
Содержание GEMS-H
Страница 1: ...GEMS H Instructions for Use ...
Страница 19: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 19 of 79 4 6 Marking plate ...
Страница 38: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 38 of 79 Uncheck Unit US in order to use SI Units instead of US Units ...
Страница 52: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 52 of 79 6 7 3 3 Example of parameter control ...
Страница 54: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 54 of 79 Error message examples ...
Страница 64: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 64 of 79 ...
Страница 67: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 67 of 79 ...