background image

21

Tel: 886.909 602 109 Email: [email protected]

www.salukitec.com

S3101A/B provides two measurement formats including DTF return loss and DTF VSWR. As for their difference, we will not explain
here again for it is similar to the difference between return loss and VSWR. We will make a detailed introduction for measurement
steps on cable length and return loss at each point, through taking the cable with length of 25m, velocity factor of 0.76 and cable loss
of 0.26dB/m.
1. Press

Meas

to enter the measurement menu;

2. Select [DTF Return Loss] menu to set the current measurement format to “DTF Return Loss”.
3. Press

Freq

to enter the frequency menu; select [Start Freq] menu, then use the numeric keys, knob or arrow keys to input

the start frequency value and select the corresponding unit menu to complete the input; the frequency in this example is set to
1GHz.

4. Similar to step 3, set the stop frequency needed by the measurement; the frequency in this example is set to 2GHz.
5. Press

Sweep

to enter Sweep menu; select [Point] to set the required sweep point, which is set to 401 here.

6. Select [Start Distance] menu, input appropriate measurement start distance by the numeric keys, knob or arrow keys, and then

select the unit menu [Meter] to finish the setting.

7. Similar to step 6, set the measurement stop distance. Once the test points and frequency are set, the current maximal test

distance of Analyzer is determined (it is displayed in the lower left corner of the screen) too. Users can adjust stop distance
based on this value and estimated position of measured fault points. If the current maximum distance is less than the
measurement distance, sweep span and points can be properly adjusted. Detailed explanation can be found in “DTF
Measurement Distance & Resolution” part in Section 3.5 of this chapter.

8. According to Section 3.1 in this Chapter, calibrate port 1 of Analyzer;
9. Press

Freq

->[Velo Factor] to set velocity factor of DUT, which is the transmission speed and light speed of electromagnetic

wave in the cable, is between 0-1.

10. Select [Cable Loss] menu, and set cable loss of the measured cable according to the known parameters of DUT.

Velocity factor and cable loss are inherent parameters of cable, if any further questions, please check the cable instructions or
consult the cable manufacturer.

11. DTF return loss curve can be measured and obtained by connecting DUT to Analyzer and then link the mismatch load or open

circuit to the end of the cable (as fault point). DTF return loss test connection is as shown in Figure 3.10.

Содержание S3101 Series

Страница 1: ...S3101 Cable Antenna Analyzer Quick Start Guide Saluki Technology Inc...

Страница 2: ...and User Manual 1 PC 8 Certificate of Calibration 1 PC Options of the S3101 cable antenna analyzer Option No Item S3101 01 USB Power Measurement Software S3101 02 Backup Rechargeable Li ion Battery S...

Страница 3: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 3 Option No Item S3101 15 Low Loss Test Cable N JJ 80cm S3101 16 Functional Soft Case S3101 18 Hard Case Carrying Case...

Страница 4: ...eets the indicator requirements of the manual at the time of delivery Calibration and measurement are completed by the measuring organization with qualifications specified by the state and relevant da...

Страница 5: ...886 909 602 109 Email sales salukitec com www salukitec com 5 Contacts Service Tel 886 909 602 109 Website www salukitec com Email sales salukitec com Address No 367 Fuxing N Road Taipei 105 Taiwan R...

Страница 6: ...Operating Interface 12 2 5 Top Interface 12 2 5 1 Power Interface 12 2 5 2 Digital Interface 13 2 5 3 Test Port 13 2 5 4 Instrument Symbol 14 2 6 Battery 14 2 7 Turn On Off 14 3 Typical Application 1...

Страница 7: ...reader can have a more systematic understanding on the Analyzer and quickly grasp some basic operations For your proficiency of the instrument please read the guide carefully and operate properly upon...

Страница 8: ...e If disassemble it at your own discretion it may cause various consequences such as degradation of electromagnetic shielding performance and damage to internal parts thus affecting product reliabilit...

Страница 9: ...d wetted and shall keep away from children Rechargeable batteries can be reused and shall be placed in a suitable container to avoid short circuit The heavy metals such as nickel and chromium containe...

Страница 10: ...the menu button on the touch screen is expressed with XXX XXX is menu name 2 3 Front Panel This section provides a detailed introduction of the front panel of S3101A B Figure 2 1 is the front panel of...

Страница 11: ...de it is used to return the Analyzer to local function Mean increase and decrease respectively and used to control step or select current options Backspace Key Used to delete the last entered number o...

Страница 12: ...ttime and normal testing environment In addition this instrument also supports automatic adjustment of brightness hibernation and automatic shutdown Figure 2 2 Operating Interface 2 5 Top Interface S3...

Страница 13: ...Antenna Analyzer 3 Mini USB interface used to connect an external PC and then related tools and software can be operated by PC to perform program control and data transmission to S3101 Cable Antenna...

Страница 14: ...g stand by batteries of the same model with the original one is recommended to avoid test interrupt owing to low battery if long time field test is performed S3101 Cable Antenna Analyzer is easy to in...

Страница 15: ...t and the user can observe whether the instrument is working properly In order to make the internal components achieve a stable performance indicator the S3101 should be warmed up for 15 minutes befor...

Страница 16: ...ess is as follows 1 Press Cal key click Cal Kit menu to select the calibration kit model and then click Enter 2 Click M Cal and the calibration wizard will prompt saying Please connect OPEN and start...

Страница 17: ...ease connect LOAD and start calibrating by pressing corresponding soft key 5 Disconnect the short standard with the calibration port connect the load and press Load key then the screen will prompt LOA...

Страница 18: ...re 1 Press MEAS to enter the measurement menu 2 Select Return Loss menu to set the current measurement format of instrument as Return Loss as shown in Figure 3 5 Figure 3 5 Selection of Measurement Fo...

Страница 19: ...epresented by average value During the test the cable loss is usually measured by the following connection as shown in Figure 3 7 Figure 3 7 Cable Loss Measurement Connections Now this part takes the...

Страница 20: ...Marker 1 onto the peak position of the cable loss curve Select Marker 1 2 3 4 5 6 menu again the menu becomes into Marker 1 2 3 4 5 6 at this time and then select Marker Off On menu to open Marker 2...

Страница 21: ...ric keys knob or arrow keys and then select the unit menu Meter to finish the setting 7 Similar to step 6 set the measurement stop distance Once the test points and frequency are set the current maxim...

Страница 22: ...are shown in Figure 3 11 Figure 3 11 Measurement on Length of Cable 3 3 Antenna Test S3101 series can test the return loss VSWR of general passive receiving antenna This part will introduce the measur...

Страница 23: ...hat you can see the VSWR situation of antenna at each frequency Through observation we can find that VSWR of antenna is smaller near 2 100MHz with minimum at 2 104MHz Figure 3 13 VSWR Test of Terminal...

Страница 24: ...ed in Section 3 5 we can find that the return losses of attenuator are all below 30dB Based on the test data designer can confirm attenuator index by contrasting with design index as shown in Figure 3...

Страница 25: ...tment 2 Limit line Users have their own eligibility criteria in batch test In this regard S3101A B provides Limit function for the setting of limit line making it easier to test products Thus when lar...

Страница 26: ...function is turned on Analyzer will raise a beep beep alarm through which users can judge whether the product is qualified or not Test curve in Figure 3 19 has exceeded the set limit line Figure 3 19...

Страница 27: ...rrow keys choose Recall menu or press Enter key to complete recall of cable models Thus Analyzer will automatically match parameters for the selected cable based on its model including velocity factor...

Страница 28: ...m distance will be In terms of cable loss nominal value of effective directionality of S3101A B is equal to or more than 42dB If the reflected signal energy is less than the effective directionality t...

Страница 29: ...ory repair If S3101A B fails to start up and self test please press Reset key which makes S3101A B return to a known state If the self test still fails then it might be an instrument failure please co...

Страница 30: ...r difficult to solve you can contact us by email or phone If you confirm that the instrument needs to be repaired please follow the steps below to package the instrument 1 Write a paper document descr...

Отзывы: