Tel: 886.909 602 109 Email: [email protected]
2.
Ready to Use
S3101 Cable and Antenna Analyzer is designed with exquisite appearance and comfortable operation. It has smaller volume, with
the largest overall dimension being only 205mm (H) × 295mm (W) × 70mm (D); and lighter weight, with the single unit weighing
totally about 2.5kg (including battery), which is easy to carry for field test. This Chapter will emphatically introduce the
instrument
’
s test environment, power supply, structure and battery replacement.
2.1.
Pre-operation Preparation
This section will describe the precautions before initial use of S3101A/B Cable & Antenna Analyzer. It meets the requirements of
security specified in GJB3947A-2009, please read the following safety instructions carefully to avoid damage to the instrument or
unnecessary personal injury.
To prevent damage to the instrument, avoid electric shock, fire and personal injury:
Do not open up the instrument without authorization
Do not attempt to disassemble or modify any part not described in this guide. If disassemble it at your own discretion, it may cause
various consequences such as degradation of electromagnetic shielding performance and damage to internal parts, thus affecting
product reliability. If disassemble it without authorization, even during the warranty period, we will no longer provide free
maintenance.
2.1.1.
Environmental Requirements
To ensure the service life and validity and accuracy of S3101A/B, please conduct testing under the following conditions:
Temperature range: Storage temperature range: -40
℃
to +70
℃
; operating temperature range: -10
℃
to +50
℃
Low pressure (altitude): 0 - 4600m
2.1.2.
Power Supply Requirements
S3101A/B can be powered in three ways:
1. Powered by AC power supply and adapter
AC-DC adapter provided together with the instrument must be used if adopting AC power supply. Input of adapter is AC power
supply with 100~240V, 50/60Hz.
When transported and carried by a backpack, the AC-DC adapter should not be connected with the Analyzer in order to avoid
overheating of the instrument. AC-DC adapter has a wide range of input voltage; therefore when using, it should ensure the power
supply voltage should be within the scope required in Table 2.1.
Table 2.1 Power Supply Requirements
Power Supply Parameter
Applicable Range