16
Tel: 886.909 602 109 Email: [email protected]
3.
Typical Application
S3101 Cable & Antenna Analyzer offers many measurement formats, and this Chapter will make a detailed instruction on operation
steps of several basic measurement formats by three test cases.
3.1.
Calibration
To ensure accuracy of test results, the instrument needs to be calibrated before test to eliminate the errors. S3101A/B adopts
mechanical calibration method, whose calibration process is as follows:
1) Press
【
Cal
】
key, click [Cal Kit] menu to select the calibration kit model and then click Enter;
2) Click [M Cal], and the calibration wizard will prompt saying “Please connect [OPEN], and start calibrating by pressing
corresponding soft key
!
”;
3) Connect the open standard to the calibration port, press [Open] key and then the prompt massage saying “[OPEN]
Measuring...” as shown in Figure 3.1:
Figure 3.1 Calibration Open Standard
After completing open circuit measurement, the [Open] button on the right side of the menu bar is underlined as [Open], and the
prompt massage saying “Please connect [SHORT], and start calibrating by pressing corresponding soft key!” as shown in Figure
3.2: