ON/OFF/ZERO
and
UNIT
key until ‘
SEtUP
’ is shown, that means the
scale is in SETUP mode.
2. This indictor has 19 kinds of parameters to be selected and setup by this
function.
3. During SETUP mode, press
UNIT
key to change the flashed digits, and
HOLD/PRINT
key to shift the flashed position. Press
TARE
key to confirm
and save the set data and enter next setting. Press
ON/OFF/ZERO
key to
exit this mode.
4. Display
1) P
A.B
: Item A parameters and one digit can be input.
2) P
A.BC:
Item A parameters and two digits can be input
3) P
AB.C
: Item AB parameters and one digit can be input
4) P
AB.CD
: Item AB parameters and two digits can be input
5. Detailed
setup:
5.1) P1.xy: Auto-off time xy= 00-15
00: means no auto-off function
01-15: means the scale turns off to save power after 01-15 minutes
when no variety of weight on scale and no key operation.
5.2) P2.x:
HOLD/PRINT
key function definition
x=0 means only HOLD function
x=1 means only PRINT function
x=2 means HOLD and PRINT function: short-time click the
button once for Print function, and hold the button for
several seconds for HOLD function
5.3) P3.xy: Hold function setting
Xy=0: No hold function;
Xy=1: When the scale becomes stable, it will hold the larger weight
reading;
Xy=2: When the scale becomes stable, it will automatically display and
hold the weight reading. When this weight is below 10d, HOLD
function will automatically inactivated. When the new weight is more
than 10d, the scale will hold the new weight when stable.
Xy=3-50: When the weight variety is within the setting range +3~50d,
the weight reading will remain unchangeable.
11
5.4) P4.x: RS232 mode setting
x=0: No RS232 function. It will not transmit or receive any data
although the scale is with RS232 hardware. RS232 function can be only
activated when scale is in normal weighing mode.
x=1:Press PRINT/HOLD key, the scale will output the current stable
displayed weight reading and weight unit, and not receive any data
from other equipment. The output format is as below:
<LF>< reading, minus, decimal point, weight unit><CR><EXT>
x=2: Press PRINT/HOLD key, the scale will output the data of stable
current gross, tare, net weight reading and weight unit. The format
is
as
follows:
<LF><Gross: reading, minus, decimal point, unit><CR><EXT>
<LF> <Tare: reading, decimal point, unit><CR><EXT>
<LF> <Net: reading, minus, decimal point, unit><CR><EXT>
The number of position used: weight reading ---7;
Minus
---1;
Decimal
point
---1;
Weight unit ---2 or 4;
x=3: Continuously output of the current displayed reading and unit, and it
does not receive any data. The output format is same as x=1.
x=4: Continuously output of the current gross weight, tare weight and net
weight reading data including unit, and not accept any data. The
output format is same as x=2.
X=5: When the scale is stable, it will output the current displayed weight
reading automatically one time including unit, and not accept data.
The output format is same as x=1.
X=6: When the scale is stable, it will output the current gross weight, tare
weight and net weight data including unit automatically one time, and not
accept data. The output format is same as x=2.
X=7: Bio-RS232 data output compatible to the NCI-SP1 format.
5.5) P5.x: Baud rate of RS232 communication
x=0:
1200bps
x=1:
2400bps
x=2:
4800bps
12