Access to the calibration switch may be sealed with a lead-wire combination 
on the bottom of the scale. Prior to removing the lead seal, check to see if the 
test switch is in the correct position by following step 1 below. If you are not 
allowed access, remove the lead-wire seal.

Note: You can exit the calibration procedure at any time by pressing the On/
Off 
key.

1.

To enter calibration mode, while in normal weighing mode, hold the 

Zero

 

and 

On/Off

 keys at the same time until…

The scale beeps and, if the calibration switch is in the ON position, 
the WEIGHT window shows 

CAL. oN

. If the display shows 

CAL.oFF

change the calibration switch position. See Figure 

3

The PIECE WEIGHT window shows 

Unit. 1

 or 

Unit. 0

. If 

Unit. 1 

is 

displayed, the calibration unit of measure is pounds. If 

Unit. 0 

is 

displayed, the calibration unit of measure is kilograms. 

2.

Press the 

Units

 key to toggle between pounds and kilograms. When the 

unit you want is displayed, press the 

Enter

 key to accept it…

The scale beeps and 

unLoAd

 is displayed in the PIECE WEIGHT 

window. A number representing raw A to D counts is displayed in the 
COUNT window.

3.

Remove all weight from the scale, wait for the scale to be stable (the 

Unit

 annunciator stops flashing) and press the 

Enter

 key to capture the 

zero point…

If the scale successfully captures the zero point, the scale beeps and 

LoAd

 is displayed in the PIECE WEIGHT window.

4.

Load a test weight that is between 25% and 100% of full capacity onto 
the center of the weight platter, wait for the scale to be stable (the Unit 
annunciator stops flashing) and press the 

Enter

 key to capture the 

span.

When the scale captures the span, 

InP. Ld

 is displayed in the PIECE 

WEIGHT window and 

0.000

 is displayed in the COUNT window. 

InP. 

Ld

 stands for input load weight.

5.

Remove the weight from the scale and key in the amount of weight you 
used. The weight entered will display in the count window. Press the 

Enter

 key…

When the scale records this weight, 

unLoAd

 is displayed in the 

PIECE WEIGHT window.

Содержание B 120

Страница 1: ...p n AWT35 500140 Issue AB July 2007 MODEL B 120 Counting Scale User Manual...

Страница 2: ...is intended for use in dry conditions For your protection all mains 110V or 240V equipment used out of doors or in wet or damp conditions should have adequate environmental protection We recommend a c...

Страница 3: ...20 hours Place the scale on a stable level surface Level the scale using the adjustable feet and the bubble level at the lower left of the front panel Tighten the locking ring on each foot once the sc...

Страница 4: ...rminal and the black wire to the black battery terminal Place battery in the compartment with the padding placed as shown and replace the battery cover and tighten the screw Note To improve battery li...

Страница 5: ...This lights when a tare is in effect Weight displayed is net weight kg This lights when the unit of measure in the WEIGHT display is kilograms lb This lights when the unit of measure in the WEIGHT dis...

Страница 6: ...100 of capacity an empty container or weight not required in the calculation of an average piece weight Use of tare will reduce the scale s maximum weight capacity by the tare value stored Accu Used t...

Страница 7: ...objects in a container follow these steps 1 Turn the scale on and zero the display if necessary by pressing the Zero key Be sure the scale is displaying weight in your preferred unit of measure Chang...

Страница 8: ...lays the number entered If the piece weight window shows SPL PCS press the CLEAR key and increase your sample weight Counting by Sampling Follow these steps to count items using the sampling method Fo...

Страница 9: ...ntering a Known Piece Weight You can enter a known piece weight and begin counting without sampling This can also be used with the tare function active After you have set the piece weight place the it...

Страница 10: ...e scale on and zero the display if necessary using the Zero key Be sure the scale is displaying weight in your preferred unit of measure Change it using the Units key if necessary 2 Use sampling or en...

Страница 11: ...nt is displayed press the Accu key to accumulate the count to memory The count is added into memory and temporarily Add is displayed in the WEIGHT window the number of accumulations is displayed in th...

Страница 12: ...he COUNT window will alternately flash H 1234 and L 5678 Note Remove the counted items from the scale and let the scale return to stable condition before you can count more items and perform a success...

Страница 13: ...ograms When the unit you want is displayed press the Enter key to accept it The scale beeps and unLoAd is displayed in the PIECE WEIGHT window A number representing raw A to D counts is displayed in t...

Страница 14: ...WEIGHT window the word codE is displayed in the PIECE WEIGHT window and the A to D counts code is displayed in the COUNT window Vol stands for voltage and X X is the current voltage of the battery If...

Страница 15: ...ly 2 Key in the number of minutes of scale inactivity that will trigger the auto off function and press the Enter key to accept this value The scale returns to normal mode 8 0 Setting LED Brightness Y...

Страница 16: ...Err There is an error in calibration maybe the loaded weight is too small or the input weight is not correct Ad Input signal of ADC is over max range Ad _ _ _ _ Input signal of ADC is below min range...

Страница 17: ...s below 5 5V the scale beeper will sound and work for approximately two minutes before powering off When the scale is powered by the AC adapter the AC annunciator will turn on When the battery is char...

Страница 18: ...16...

Страница 19: ...0 444 6132 Fax 44 0 870 010 2241 Email sales salterbrecknell co uk Web Site www averyweigh tronix com USA Salter Brecknell Weighing Products USA 1000 Armstrong Drive Fairmont MN 56031 Toll free 800 63...

Отзывы: