TROUBLESHOOTING CHART
DISPLAY CODE
SOURCE OF ERROR
SOLUTION
0
Mag. Empty manual load
Insert slide on chute
1
Mag. Empty station 1
Reload slide magazine
2
Mag. Empty station 2
Reload slide magazine
3
Mag. Empty station 3
Reload slide magazine
Tray load led flashes
Unload station tray needs replacement
Power led flashes
System in sleep mode
00
Print head clean in progress
Ink empty LED on
Ink cartridge needs replacement
11
Flash power supply over heated
Call service
12
No heating in flash unit
Check heat gun
13
Flash light reached lifetime
Replace flash light
14
Maintenance indicator
Call service
15
Manual Cleaning print head demanded
Perform print head cleaning procedure
20
The magazine is mechanically blocked
Remove the cause of the blockage
21
Faulty ejection of a slide or slide. Magazine is
blocked
Remove the slide
22
The transfer of the slide from the loader chute to
the print carriage failed. Horizontal motor is
incorrectly positioned or mechanically blocked
Remove the slide
23
Slide is stuck in the delivery chute
Remove the slide
32
The horizontal drive is mechanically blocked
Remove the slide
33
The vertical drive is mechanically blocked
Remove the slide
34
The rotational motion is mechanically blocked
Remove the slide
35
Faulty clamping of a slide. A slide left the
delivery chute but did not reach the print sledge
Remove the slide from the print sledge
36
The slide did not leave the print sledge or was
still located in the clamp during an initialization
Remove the slide from the print sledge
37
The ink print head is too hot. Ambient
temperature too high or electronics defective
Switch the instrument to the off position and let
cool. The instrument will remain disabled until
the print head cools to a regular temperature.
Check the ambient temperature is within
normal ranges
38
No or incorrect voltage at ink print head
Contact your technical support representative
41
Flash standby state not reached within the
prescribed time. Charging electronics defective.
Contact your technical support representative
42
Flash does not occur at all or is too short. Curing
lamp dirty or defective.
Check whether flash takes place-to do so,
observe the stray light coming from under the
cover. Never open the cover to check whether
the flash takes place!
No flash, install a new curing lamp
43
The temperature in the drying station is not
within the permissible temperature range
Carry out a RESET
Содержание Tissue-Tek Autowrite
Страница 1: ...3 4 2011 Tissue Tek Autowrite Slide Printer ...
Страница 18: ...1 14 Revised 3 04 11 INTRODUCTION ...
Страница 32: ...2 14 Revised 3 04 11 INSTALLATION SETUP ...
Страница 43: ...Revised 3 04 11 3 11 Section 3 ...
Страница 62: ......
Страница 68: ...8 2 Revised 3 04 11 SERVICE AND REPLACEMENT PARTS ...