7.2
Revised 3/04/11
TROUBLESHOOTING
Replacing the Curing Lamp
Code 13 appears on the instrument display when the
curing lamp reaches the limit of its rated service life. When
this code is issued, the lamp must be changed. See
installation/removal procedure on page 6.2.
CAUTION: Before replacing the curing lamp, switch
the printer off and unplug from the electrical outlet.
Once the lamp has been replaced, plug the instrument
into the electrical outlet and turn the power switch to the
on position.
Press and hold the
[ERROR]
key for approximately 10
seconds to activate the off-line mode. Code 13 will
disappear from the display.
Power Failure
If the instrument suddenly loses power:
Determine if there is a general power failure
Check to see if the power cord is inserted correctly
into the outlet and if there is power to the outlet.
Check the power switch to determine if it is turned to
the ON "I" position. It is possible that the primary fuse
has reacted in which case the power switch will be in
the OFF "O" position.
Determine if one of the four secondary fuses is
defective.
The following malfunctions/failures may be caused by the
indicated fuse:
Instrument not functioning
Fuse F1
Display not working
Fuse F1
Instrument not working at normal speed
(printing a slide slows down to approximately 8 seconds
even after the warm up phase has been completed)
Fuse F2
Replacing the Secondary Fuses
CAUTION: Prior to exchanging the fuse, turn the
instrument off and unplug from the power outlet.
NOTE:
Defective fuses may be replaced only with the
replacement fuses supplied with the instrument or
purchased from Sakura.
1. Insert a screwdriver into the slot on the fuse holder
(Figure 7-B). In one continuous motion, push slightly
inwards and rotate the screwdriver ¼ turn to the left.
Figure 7-B
2. Take out the released fuse holder
3. Remove the defective fuse from the fuse holder
(Figure 7-C) and insert the replacement fuse.
Figure 7-C
4. Re-insert the fuse holder into the instrument housing.
5. Insert the screwdriver into the slot on the fuse holder.
In one continuous motion, push slightly inwards and
rotate the screwdriver ¼ turn to the right.
Содержание Tissue-Tek Autowrite
Страница 1: ...3 4 2011 Tissue Tek Autowrite Slide Printer ...
Страница 18: ...1 14 Revised 3 04 11 INTRODUCTION ...
Страница 32: ...2 14 Revised 3 04 11 INSTALLATION SETUP ...
Страница 43: ...Revised 3 04 11 3 11 Section 3 ...
Страница 62: ......
Страница 68: ...8 2 Revised 3 04 11 SERVICE AND REPLACEMENT PARTS ...