Revised 3/04/11
6.7
Section 6
7. Reinsert the location plate (Figure 6-S).
Figure 6-S
NOTE:
The location plate must be completely dry before
reinserting.
8. Press any key to acknowledge the end of the cleaning
procedure. The print head will move back to the
parked position and the message will move back to the
parked position and the message
“15” will disappear
from the display. If no key is pressed to acknowledge
the end of the cleaning procedure, the print head will
be closed automatically after a few minutes to prevent
it from drying out.
Transportation/Storage
When preparing to transport or store the AutoWrite
Slide Printer, the ink cartridge should be replaced with a
transport cartridge. Follow the instructions outlined in
Removing and Replacing the Ink Cartridge
on page 6.4.