25
3410330_GB
CAUTION:
If the equipment is operated with feed water that does not have
the characteristics listed in the above table, malfunctions and damage may
occur for which the Manufacturer accepts no liability.
CAUTION: The Manufacturer's warranty on parts damaged by a water supply that
does not correspond to the characteristics listed in the above table may be
invalidated.
SUGGESTIONS ON FILTRATION SYSTEMS
In the event that the water parameters do not comply with the characteristics in the previous
table, you can use the suggestions below when choosing the filtration system to be installed,
depending on the parameter that needs to be corrected. These suggestions should then be
considered as indicative of certain situations and not exhaustive of all cases that may arise.
However, it is always the responsibility of the Owner of the equipment to ensure that the
feed water corresponds to the parameters required in the table above, even after installation
of the treatment system.
It is advisable to contact personnel or companies specialised in water analysis and treatment
to ensure that the feed water meets the required characteristics.
The parameters that are best treated by the corresponding system are then marked in bold
in the table.
Parameters
Measured value
Suggested
type
of
treatment
Carbonate hardness
Chlorine
Chlorides
Sulphates
>9°f
<0.1 ppm
<10 ppm
<30 ppm
Softening
Carbonate hardness
Chlorine
Chlorides
Sulphates
<9°f
0.1 - 1 ppm
<10 ppm
<30 ppm
Activated carbon filter
Carbonate hardness
Chlorine
Chlorides
Sulphates
>3°f
<1 ppm
10 - 100 ppm
30 - 100 ppm
Special resin filter or
reverse osmosis treatment
Carbonate hardness
Chlorine
Chlorides
Sulphates
>3°f
<10 ppm
>100 ppm
>100 ppm
Reverse osmosis
(tab 9)