Sage Technologies
32
Cut off damaged O-rings with an X-acto knife. Then replace the one clear and two black O-rings
on the shuttle. The o-rings can be loaded onto the shuttle valve with the maintenance tool: separate the
maintenance tool into two pieces, then screw the tapered piece of the tool into the end of the shuttle valve.
The o-rings can then be pushed onto the end of the tool and slid onto the shuttle valve in order. The white
o-ring is in the middle, between the two black o-rings. (These o-rings are supplied in the gun maintenance
kit, which comes with the gun at time of purchase.)
When replacing any of the o-rings, wash the shuttle valve passage and grease the shuttle valve
and the valve passage liberally with silicone lubricant. Reinstall the shuttle valve with the long end of the
shuttle valve to the “Fire” side of the gas gun.
The inside of the gun may be cleaned by spraying it with petroleum solvent and/or by wiping the
surface clean. Grease threads and O-rings liberally with lubricant before reassembling.
Replace the bonnet top on gun, aligning the microphone holes in the top and bottom. Replace and
tighten screws as you would a flange -- tighten in opposing fashion by tightening one screw loosely, then
the one across from it (not beside it). After all screws are tightened loosely, tighten with ball-end socket
wrench, again in opposing fashion.
Before reinstalling the microphone, check the o-rings for damage, and replace if necessary.
Lubricate the o-rings with silicone lubricant before sliding the microphone back into the gun and securing
it with the microphone bolt.