Acoustic Fluid Logger and Pressure Pulse Gas Gun
33
For maintenance on the fill valve, remove it with the thick, notched end of the gun maintenance
tool, turning counter clockwise. An o-ring and the valve core will be exposed under the fill valve.
Remove the valve core with the tapered, notched end of the gun maintenance tool, turning
counterclockwise. Replace the old valve core with a new one, and hand tighten, using the maintenance
tool. Replace the o-ring, if necessary. Replace the fill valve and tighten securely.