If not, it will mean the sensitivity setting can not
be improved upon and cancellation will be
ineffective.
v) Select YES
vi) The product will require three further
passes, after the first
Will be displayed. After the second
will be displayed. After the third
will be displayed. The final message will
be
vii)If at any stage an error has occurred or
the PROG/EXIT key has been pressed,
the following will be displayed
and the settings from the first part of
Product adjustment will be used with PSC
not working.
viii) The letter 'P' in 'PHASE' will flash
(Firmware Version 2.15 or greater)
indicating PSC is working.
IMPORTANT
To achieve optimum performance and prevent
false detections, the conveyor belt MUST be
free from contamination.
The position of the pack sensor MUST detect
the pack at a regular point. If the pack has a
tag or joint on the leading edge, the pack
sensor detection point could be inconsistent
and lead to impaired operation.
Once PSC is setup and working, if required
SENS can be manually adjusted but PHASE
cannot. If PHASE is adjusted PSC will stop
working and the 'P' in PHASE will stop flashing
(Firmware Version 2.15 or greater).
VIEWING Mode
The VIEWING mode is a new access level.
The user may view all settings that are
normally set in the ENGINEERS mode but
cannot change them. The default code is
0005.
This may be changed in the ENGINEERS
access mode.
Japanese Language
Japanese has been added to the list of
languages available to the metal detector. As
with the other language options, all displays
have been translated. Printouts however, have
not, they will remain in English (2.21 and
above have Japanese printouts).
Due/Overdue Alarm Output
There has been a change to a Performance
Validation Routine setting. In QA SETUP,
Alarm if Overdue - YES/NO
has changed to,
Alarm if - - DUE/OVERDUE
An independent alarm output is available if a
Performance Check becomes either due or
overdue.
This is an open collector output, capable of
sinking a maximum current of 100 mA, at a
maximum voltage of 30 V.
(Note - reject confirmation does not have to be
selected).
Page 47B
Содержание STANDARD SERIES
Страница 11: ......
Страница 22: ...Figure 6C Cable Gland Assemblies for Power Supply Unit Page 9A...
Страница 23: ...Figure 8 Bearing Block Insulation Figure 7 Roller Shaft Insulation one end only Page 9B...
Страница 24: ...THIS PAGE IS LEFT BLANK INTENTIONALLY Page 9C...
Страница 25: ...Page 10...
Страница 26: ......
Страница 78: ...Page 60 TYPICAL METAL DETECTOR CONVEYOR...