34
Error code
Cause & Remedy
Cause: The foot controller connection is not proper or the foot controller is defective.
Remedy: Make sure that the foot controller is properly connected.
Contact the manufacturer if the remedies do not activated
Cause: The motor connection is not
proper or the motor is defective.
Remedy:
M
ake sure that motor properly
connected. Contact the
manufacturer if the remedies
do not activated.
Cause: Torque is overloaded than the setting values during the surgery.
Remedy: Press the speed control pedal to release the error.
Contact the manufacturer if the remedies do not activated.
Cause: Select motor without motor connector
(Motor 1 / Motor 2)
Remed
y: Press the motor button to be used (Motor 1 /
Motor 2). Contact the manufacturer if the
remedies do not activated.
■ Error & Remedy
: In case of stopping handpiece by not working & using error, each error code may be shown below.
1) Angle Handpeice
① Rotate the head in a tap water about 10 seconds with maximum speed to remove the remains
(blood, physiologic saline solution, etc.) and rinse.
※ Do not put tap water in the motor.
② Disconnect the angle handpiece to the motor.
③ Remove contaminants on the angle handpiece using a nylon brush after soaking for 20 minutes in Enzymatic Detergent.
④ Rinse the angle handpiece using the tap water for 3 minutes to remove Enzymatic Detergent.
⑤ Wash the angle handpiece for 10 minutes in the ultrasonic cleaner containing Enzymatic Detergent.
⑥ Rinse the angle handpiece for 3 minutes in the ultrasonic cleaner containing purified water repeatedly subjected 3 times.
(Change the purified water each time.)
⑦ Wipe using dust-absorbent after the product soaked in alcohol(70%) for 30 seconds.
2) Motor
Wipe the surface of motor with alcohol swab or paper towel after removing the dirt.
※ Pay attention to not enter the water into the motor.
■ Fuse replacement
1. Cleaning
Check the fuse in case the control box is malfunctioning.
Contact the manufacturer for replacement.
Fuse box
MAINTENANCE OF THE PRODUCT
Содержание TRAUS SSG10
Страница 1: ...www saeshin com USER MANUAL SSG10...
Страница 2: ...SSG10...
Страница 4: ...I BF IEC 60529 IPX0 IPX8 DRILL SAW SHAVER 3 10 1 hPa hPa 10 30 10 80 700 1 060 20 60 10 90 500 1 060...
Страница 6: ...AC 100 120V 220 240V 3...
Страница 8: ...LCD EMC EMC Electromagnetic Compatibility ISO 1797 1 1 2 3 4 1 2 3 4 5...
Страница 14: ...SETUP SETUP SETUP 1 2 11 LCD...
Страница 20: ...1 1 1 1 2 2 1 1 2 3 1 1 0 4 3 2 1 4 2 3 5 1 4 17...
Страница 22: ...1 2 3 O X 19 4 A B C B 1 D C 1...
Страница 26: ...23 DRILL SAW SHAVER 8 9 1 2 3 1 2 3 1 2 3 1 2 3 1 2 3 1 2 3...
Страница 36: ...33 M1 M2 M1 M2 1 10 20 3 10 3 3 70 30 2 1...
Страница 42: ...IEC 60529 TRAUS FUS10 IPX8 6 39 1 1 2 SETUP SETUP SETUP EU CE BF 1639 NON STERILE...
Страница 46: ...43 MEMO...
Страница 47: ...44 MEMO...