CAUTION

● 

 

Pay attention to use considering a patient’s safety first. 

● 

 

Check the working condition to see if there is any problem before operation.

● 

 

Stop working and verify the product if there is any signs of vibration, calescence or strange noise before or during operation.

● 

 

Contact the manufacturer if the problem is abnormal or difficult to solve.

● 

 

Prepare extra consumables before using for safety. In case blood or saline solution flows into the head of the angle, which may cause malfunctioning.

● 

 

Change Bur, Blade after the motor stops completely. Otherwise, Inside of angle could be damaged by the Bur, Blade, and it will cause an accident.

● 

 

Please sterilize and lubricate products immediately after use following the manual. Otherwise, the product can be damaged due to coagulation of blood

     or remains from operation.

● 

 

The control box and foot controller are not steam sterilization objects.

● 

 

The input power is available at AC 100-120 V / 220-240 V.

     (Check input voltage before connecting the AC power.)

● 

 

Dispose the irrigation tube after use as medical waste.

● 

 

Contact to manufacturer if it shows the fault display on the LCD. It may cause an accident.

● 

 

Do not drop from high or give a shock on the product. It can damage the product.

● 

 

Do not give any damage to irrigation tube, it will not work normally due to the tube is broken or taken off.

● 

 

Do not insert oil into the motor. It will damage bearing or cause calescence.

● 

 

This product is designed to be used indoors. Do not use outdoors.

● 

 

Do not place or use the product in humid or dusty places.

● 

 

Use products and consumables approved by manufacturer. Otherwise there might be a risk of malfunction or accident.

● 

 

Replace the motor cord immediately when it is worn out or damaged to prevent electric shock for user or patient.

● 

 

Product specification and design can be changed for improvement of the product.

● 

 

User is responsible for operation and maintenance of the product.

● 

 

The company did not sterilize the product. Please be sure to sterilize before operation. (User sterilization)

● 

 

US Federal law restricts this device to sale by or on the order of a dentist.

 

USER GUIDELINES

4

Содержание TRAUS SSG10

Страница 1: ...www saeshin com USER MANUAL SSG10...

Страница 2: ...SSG10...

Страница 3: ...AGES PRODUCT COMPONENTS PART NAME OF CONTROL BOX BUTTONS ON OPERATION PANEL LCD WINDOW BUTTONS ON THE FOOT CONTROLLER CONNECTION OF EACH PART OPERATION PROCESS MAINTENANCE OF THE PRODUCT ATTENTION PRO...

Страница 4: ...I BF IEC 60529 IPX0 IPX8 DRILL SAW SHAVER 3 10 1 hPa hPa 10 30 10 80 700 1 060 20 60 10 90 500 1 060...

Страница 5: ...Type BF applied part Classification according to the degree of protection against ingress of water as detailed in the current edition of IEC 60529 Control Box IPX0 Foot Controller IPX8 Not suitable p...

Страница 6: ...AC 100 120V 220 240V 3...

Страница 7: ...ut voltage before connecting the AC power Dispose the irrigation tube after use as medical waste Contact to manufacturer if it shows the fault display on the LCD It may cause an accident Do not drop f...

Страница 8: ...LCD EMC EMC Electromagnetic Compatibility ISO 1797 1 1 2 3 4 1 2 3 4 5...

Страница 9: ...power plug It may cause the danger of fire electric shock Unauthorized modification and dismantling is prohibited Do not use the solvent such as thinner or benzene for cleaning Please use an absorbent...

Страница 10: ...DRILL SHAVER SETUP Auto Calibration 1 1 40 000 rpm 1 3 120 000 rpm 1 5 200 000 rpm 10 1 4 000 rpm 20 1 2 000 rpm 32 1 1 250 rpm 1 1 1 6 N cm 1 3 1 2 N cm 1 5 1 N cm 10 1 5 30 N cm 20 1 5 80 N cm 32 1...

Страница 11: ...hat user wants with angle handpiece for accurate and safe operation The function is set up when the product is released by manufacturer And use the function when user repair or change the product PROD...

Страница 12: ...2 13 14 17 18 21 22 23 24 19 15 16 20 No No 1 DRILL 1 1 DRILL 1 1 DRILL 10 1 SAW Reciprocating SAW Sagittal SAW Oscillating SHAVER BLDC AC DRILL SAW SHAVER Y 9 12 DRILL SAW A Y 2 13 3 14 4 15 5 16 6 1...

Страница 13: ...SAW handpiece Oscillating type Option SHAVER handpiece Option BLDC motor Option Foot controller Power cord Wrench Saline solution Suction pump set DRILL bur SAW blade SHAVER blade are not included in...

Страница 14: ...SETUP SETUP SETUP 1 2 11 LCD...

Страница 15: ...ain Power Switch Irrigation pump Motor connector Operation panel Mode button Program button Memory button Setup button Torque button Setup Setting button Speed button Setup Item shifting button Motor...

Страница 16: ...3 LED LED M2 DRILL 5 90 m min 20 0 5 1 1 0 1 2 3 4 5 0 1 35 ml min 25 2 50 ml min 25 3 65 ml min 20 4 80 ml min 20 5 90 ml min 20 1 1 1 3 1 5 10 1 20 1 32 1 2 2 SETUP SETUP SETUP SETUP DRILL DRILL SET...

Страница 17: ...n 20 Motor1 button This is used for selecting the motor 1 0 1 2 3 4 5 Stage 0 STOP Stage 1 35 ml min 25 Stage 2 50 ml min 25 Stage 3 65 ml min 20 Stage 4 80 ml min 20 Stage 5 90 ml min 20 1 1 1 3 1 5...

Страница 18: ...1 0 5 0 5 0 5 2 P1 P9 9 1 3 DRILL SAW 4 DRILL SHAVER 5 6 LED 7 DRILL SHAVER DRILL E3 SHAVER 8 9 DRILL 10 M2 DRILL 15 1 2 3 6 11 10 11 12 5 7 4 8 9 11 1 2 12 LCD...

Страница 19: ...h optic angle handpiece When press the speed control pedal on the foot controller after operating the optic function LED is turned on 7 Torque Frequency display DRILL SHAVER mode Torque display DRILL...

Страница 20: ...1 1 1 1 2 2 1 1 2 3 1 1 0 4 3 2 1 4 2 3 5 1 4 17...

Страница 21: ...tton for selecting the operation direction Press the button shortly change the rotary direction Press the button above 1 second switch the motor motor 1 motor 2 3 Irrigation button Use the button for...

Страница 22: ...1 2 3 O X 19 4 A B C B 1 D C 1...

Страница 23: ...ntrol box connector lock If you do not use it for long period please open pump roller cover Otherwise the irrigation tube might be damaged Open the pump roller cover upward Close the irrigation pump c...

Страница 24: ...21 5 SHAVER SHAVER 6 Bottle 7 DRILL SAW SHAVER TRAUS CST10XX TRAUS CRT14XX DRILL DRILL TRAUS TST10XX TRAUS TRT12NRC TRAUS TRT12NOS TRAUS TRT12NSG TRAUS TRT12NSH SAW...

Страница 25: ...n insert the irrigation tube needle to the pack in place with the pack downward 7 Irrigation nozzle connection Insert the irrigation tube to the end of spray nozzle pipe DRILL handpiece SAW handpiece...

Страница 26: ...23 DRILL SAW SHAVER 8 9 1 2 3 1 2 3 1 2 3 1 2 3 1 2 3 1 2 3...

Страница 27: ...dle clockwise to open the chuck Handle 2 Insert the bur into the chuck completely 3 Turn the handle counterclockwise to close the chuck 1 Turn the wrench counterclockwise to release the nose tip Nose...

Страница 28: ...l min 20 80 ml min 20 90 ml min 20 1 1 300 40 000rpm 1 3 900 120 000rpm 1 5 1 500 200 000rpm 10 1 30 40 000rpm 20 1 15 2 000rpm 32 1 10 1 250rpm 7 9 M2 DRILL 8 1 1 1 6 N cm 1 3 1 2N cm 1 5 1N cm 10 1...

Страница 29: ...rigation stop 50 ml min 25 65 ml min 20 80 ml min 20 90 ml min 20 1 1 300 40 000rpm 1 3 900 120 000rpm 1 5 1 500 200 000rpm 10 1 30 40 000rpm 20 1 15 2 000rpm 32 1 10 1 250rpm 7 Press the torque butto...

Страница 30: ...27 SAW 1 2 SAW 3 4 5 000 12 500 cycle min 5 35 ml min 25 50 ml min 25 65 ml min 20 80 ml min 20 90 ml min 20 SAW SAW SAW SAW SAW 3 1 10 N cm...

Страница 31: ...y pressing the speed control pedal on the foot controller 35 ml min 25 Irrigation stop 50 ml min 25 65 ml min 20 80 ml min 20 90 ml min 20 Before operating product please check the connection of irrig...

Страница 32: ...l min 25 65 ml min 20 80 ml min 20 90 ml min 20 SHAVER SHAVER SHAVER SHAVER 4 5 Hz 300 1 000rpm 0 5Hz 1 100 1 800rpm 1 0Hz 1 900 2 600rpm 1 5Hz 2 700 3 400rpm 2 0Hz 3 500 4 200rpm 2 5Hz 4 300 5 000rpm...

Страница 33: ...od of Irrigation tube Refer to the connection of each part If you use the SHAVER angle make sure that the mode set as SHAVER before use it If you do not set the SHAVER mode there is a risk of an accid...

Страница 34: ...10 N cm Hz 1 1 3 1 X X 0 0 0 P1 P1 P1 1 1 9 P9 1 P1 SETUP 1 SETUP SETUP SETUP 4 2 3 SAVING 2 AUTO CALIBRATION Calibration RPM Calibration Calibration Completed NO SETUP YES SETUP 5 10 RESET RESET Com...

Страница 35: ...program is memorized with beep sound 2 AUTO CALIBRATION After connecting the angle to the motor hold the motor by hand tightly Step the foot controller fully in order to start the calibration Calibrat...

Страница 36: ...33 M1 M2 M1 M2 1 10 20 3 10 3 3 70 30 2 1...

Страница 37: ...Angle Handpeice Rotate the head in a tap water about 10 seconds with maximum speed to remove the remains blood physiologic saline solution etc and rinse Do not put tap water in the motor Disconnect th...

Страница 38: ...35 2 3 DRILL TRAUS TST10XX TRAUS CST10XX TRAUS CRT14XX 1 3 SAW TRAUS TRT12NRC TRAUS TRT12NOS TRAUS TRT12NSG SHAVER TRAUS TRT12NSH 132 4 30 250 DRILL SAW SHAVER 1 2 132 4 30 132...

Страница 39: ...motor and motor cord when sterilizing Put the sterile bag in the sterilizer and sterilization for 4 minutes with 132 and let it dry for 30 minutes This product is designed to withstand 250 times of r...

Страница 40: ...60 Hz 5A 2A 68 VA Max 90 min 20 CA bur 2 35 ISO 3964 TRAUS TST10XX 1 1 TRAUS CST10XX 1 1 TRAUS CRT14XX 10 1 TRAUS MBP20SX 40 000 rpm TRAUS CST10XX TRAUS CRT14XX TRAUS TST10XX TRAUS TRT12NSH 4 1 ISO 3...

Страница 41: ...f time 1 Turn off and unplug the product Check the cord regularly and replace it if it has any damage 2 Clean and sterilize 3 Keep the product in proper place 1 Control box 2 Motor 3 DRILL handpiece 4...

Страница 42: ...IEC 60529 TRAUS FUS10 IPX8 6 39 1 1 2 SETUP SETUP SETUP EU CE BF 1639 NON STERILE...

Страница 43: ...e equipment into an outlet on a circuit different from that to which the other device s are connected Consult the manufacturer or filed service technician for help Electromagnetic Immunity Compliance...

Страница 44: ...icable to the frequency of the transmitter Recommended separation distance d 1 2 P d 1 2 P 80MHz to 800MHz d 2 3 P 800MHz to 2 5GHz Where P is the maximum output power rating of the transmitter in wat...

Страница 45: ...lied by manufacturer on demand 2 We will send the translated draft manual with English version to the import company to check it 3 After we get the feedback from the import company we will offer the c...

Страница 46: ...43 MEMO...

Страница 47: ...44 MEMO...

Страница 48: ...53 580 0999 1 52 5 Tel 053 587 2345 Fax 053 580 0918 www saeshin com traus saeshin com SOM S071S KE REV 4 2020 03 01 MERIDIUS MEDICAL EUROPE LIMITED Unit 3D North Point House North point Business Par...

Отзывы: