<BLDC motor designed by SAESHIN’s own technology and know>
- Provide the optimized performance by BLDC motor (0~40,000rpm) and Angle handpiece of several kinds.
<D
RILL mode>
* Maximum speed
<SAW mode>
* Maximum speed: 12,500 cycle/min
* Torque: 10 N·cm
* Gear ratio: 3:1
<SHAVER mode>
* Maximum speed: 6,000 rpm (Forward, Reverse motion)
* Frequency (F/R repetition motion):
<Program memory function>
- 9 programmable memories for setting Gear ratio, Speed, Torque, Rotating direction, Irrigation pump and Optic.
<Automatic overload protection function>
- If the load on the bur is higher than set torque, the motor will stop automatically and an Error message will be displayed.
- To release the overload function, press the foot controller once.
<Realtime RPM view system>
- Actual RPM value is displayed on the screen. It makes user can do more delicate operation.
<Motor Auto-calibration function>
- Motor auto-calibration function is that controller check the angle handpiece status to reconcile speed and torque that user wants with
angle handpiece for accurate and safe operation.
- The function is set up when the product is released by manufacturer. And use the function when user repair or change the product.
PRODUCT FEATURES & ADVANTAGES
Gear ratio
Torque
1:1
1~6 N·cm
1:3
1~2 N·cm
1:5
1 N·cm
10:1
5~30 N·cm
20:1
5~80 N·cm
32:1
5~80 N·cm
- The free-running speed of the handpieces shall be in accordance with the manufacturer’s instructions at a tolerance of
±10 % as specified. (Refer to ISO 14457)
Gear ratio
RPM
1:1
40,000 rpm
1:3
120,000 rpm
1:5
200,000 rpm
10:1
4,000 rpm
20:1
2,000 rpm
32:1
1,250 rpm
* Torque
* Gear ratio: 1:1, 1:3, 1:5, 10:1, 20:1, 32:1
* LED: 25,000 lx
RPM
Frequency
300~1,000 rpm
0.5 Hz
1,100~1,800 rpm
1.0 Hz
1,900~2,600 rpm
1.5 Hz
2,700~3,400 rpm
2.0 Hz
3,500~4,200 rpm
2.5 Hz
4,300~5,000 rpm
3.0 Hz
<Beep sound function> (DRILL, SHAVER mode)
- When using reverse direction, ring the beep sound for protecting user confusion.
<Ergonomic foot controller>
- The foot controller is designed ergonomically to control all the functions. And the foot controller provides high convenience.
<Setting with SETUP mode>
- Volume control, Time set, Program reset, Auto Calibration is available.
8
Содержание TRAUS SSG10
Страница 1: ...www saeshin com USER MANUAL SSG10...
Страница 2: ...SSG10...
Страница 4: ...I BF IEC 60529 IPX0 IPX8 DRILL SAW SHAVER 3 10 1 hPa hPa 10 30 10 80 700 1 060 20 60 10 90 500 1 060...
Страница 6: ...AC 100 120V 220 240V 3...
Страница 8: ...LCD EMC EMC Electromagnetic Compatibility ISO 1797 1 1 2 3 4 1 2 3 4 5...
Страница 14: ...SETUP SETUP SETUP 1 2 11 LCD...
Страница 20: ...1 1 1 1 2 2 1 1 2 3 1 1 0 4 3 2 1 4 2 3 5 1 4 17...
Страница 22: ...1 2 3 O X 19 4 A B C B 1 D C 1...
Страница 26: ...23 DRILL SAW SHAVER 8 9 1 2 3 1 2 3 1 2 3 1 2 3 1 2 3 1 2 3...
Страница 36: ...33 M1 M2 M1 M2 1 10 20 3 10 3 3 70 30 2 1...
Страница 42: ...IEC 60529 TRAUS FUS10 IPX8 6 39 1 1 2 SETUP SETUP SETUP EU CE BF 1639 NON STERILE...
Страница 46: ...43 MEMO...
Страница 47: ...44 MEMO...