23
Recommended separation distances between portable and mobile RF
communications equipment and the TRAUS SIP20
The TRAUS SIP20 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the TRAUS SIP20 can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the TRAUS SIP20 as recommended
below, according to the maximum output power of the communications equipment
Rated maximum output
Power of transmitter
(W)
For transmitters rated at a maximum output power not listed above, the recommended separation distance ‘d’ in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and persons.
150 kHz ~ 80 MHz
d = 1.2 √P
Separation distance according to frequency of transmitter (m)
800 MHz ~ 2.7 GHz
d = 2.3 √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
80 MHz ~ 800 MHz
d = 1.2 √P
Cables and accessories
Maximum length
Shield
Complies with
Piezo handpiece cord
Foot controller cord
AC power cord
1.7 M
1.7 M
1.7 M
Unshielded
Unshielded
Unshielded
RF emissions, CISPR11
Class A / Group 1
Harmonic emissions
IEC 61000-3-2
EN 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3
EN 61000-3-3
Electrostatic discharge (ESD)
IEC 61000-4-2
EN 61000-4-2
Electrical fast transient / burst
IEC 61000-4-4
EN 61000-4-4
Surge
IEC 61000-4-5
EN 61000-4-5
Voltage dips, short interruptions
and voltage variations on
power supply input lines
IEC 61000-4-11
EN 61000-4-11
Power frequency (50/60Hz)
magnetic field
IEC 61000-4-8
EN 61000-4-8
Conducted RF
IEC 61000-4-6
EN 61000-4-6
Radiated RF
IEC 61000-4-3
EN 61000-4-3
Description of BASIC SAFETY and ESSENTIAL PERFORMANCE
The TRAUS SUS20 is intended for use in the electromagnetic environment specified above.
The following essential performance of TRAUS SUS20 should be running in such an environment.
1) Irrigation volume should be over the 35 ㎖/min ±20% during handpiece is operated.
Содержание TRAUS SIP20
Страница 1: ...USER MANUAL www saeshin com SIP20...