14
OPERATION PROCESS
1. Press the Gear Ratio Button ( ) to select the gear ratio to use.
2. Press the Rotation Direction Button ( ) on the control box or Rotation Direction Button ( )
on the foot controller to select the rotation direction. Rotation direction is forward and reverse.
During the reverse operation, it beeps continuously.
(beeps let the user know the operation direction.)
3. Press the Irrigation Button ( ) on the control box or Irrigation Button ( ) on the foot
controller to select the irrigation volume. Irrigation volume per each stage refers to the below.
4. Press the Menu Button ( ) to go to the RPM Menu. Set the desired RPM value using and
button.
Keep pressing speed button to change rapidly. The setting range of rpm is different from gear ratio.
5. Press the Menu Button ( ) to go to the Torque Menu. Set the desired torque value using
and button.
The setting range of torque is different from gear ratio.
■
Program Setting
1. Press the Menu Button ( ) to go to program number menu.
2. Press the Control Button ( , ) to select the desired program number.
3. After setting the desired value by referring to the operation method, keep pressing OK Button ( )
for 3 seconds.
4. The ‘SAVING’ is displayed on LCD, program is memorized with beep sound.
Gear Ratio
1
:
1
1
:2
1
:5
1
6:
1
20:
1
32:
1
RPM
1
,000
~
4
0,000 rpm
2,000
~
8
0,000 rpm
5,000
~
200,000 rpm
30
~
2,500 rpm
1
5
~
2,000 rpm
10 ~
1
,250 rpm
Gear Ratio
1
:
1
1
:2
1
:5
1
6:
1
20:
1
32:
1
Tor
q
ue
-.-
5
~
50 N·cm
5
~8
0 N·cm
5
~8
0 N·cm
- Before operating product, please check the connection of irrigation tube on the angle
handpiece. If irrigation tube isn’t connected and delivered to bur, there is risk of tissue necrosis
or bone damage. Because of the frictional heat between bur and operation site.
(Installation method of Irrigation tube: Refer to the connection of each part.)
6. Press the Optic Button ( ) to use optic function.
7. Use it by pressing the speed control pedal on the foot controller.
8. After the operation is finished, remove the foot from the speed control pedal, stop the product,
and turn off the product
Irrigation Stop 35 mℓ/min ± 25% 50 mℓ/min ± 25% 65 mℓ/min ± 20% 80 mℓ/min ± 20% 90 mℓ/min ± 20%
-.-
-.-
Содержание TRAUS SIP20
Страница 1: ...USER MANUAL www saeshin com SIP20...