![RWD D180P Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/rwd/d180p/d180p_user-manual_1503442009.webp)
separate collecting system for electronic waste. This product must be
disposed of within the waste management regulations.
- Warning!
Parts and components that are susceptible to electrostatic discharge
(ESD) are included
–
Asphyxiation Hazard Warning.
High concentrations of CO2 can displace oxygen and cause
asphyxiation!
–
Equipment being maintained or serviced must be turned off and
locked off to prevent possible injury.
2.2 Safety instructions
Plug in the power cord to prevent improper power contact;
Pay attention to shockproof, waterproof, moisture proof, pressure proof and
fireproof measures;
During moving and handling equipment, pay attention to handling strength to
prevent damages to equipment or falling of equipment;
The CO
2
Incubator should be managed by specific personnel. It is strictly
forbidden for untrained personnel to operate the equipment. Please use a
reliable grounding device.
The responsibility for equipment failure arising from any improper cleaning,
maintenance and operation of the equipment should be borne by the user.
If the equipment host is disassembled without the authorization of RWD, RWD
will no longer fulfill its commitments for quality assurance and technical
maintenance services to the equipment. For any technical problems please
contact the authorized personnel or RWD for supports.
2.3 Precautions for operation
The
Incubator should be regularly cleaned and disinfected with
disinfectant. Its stainless steel shelves can be removed for cleaning and
disinfection to avoid failure of the experiment due to contamination by other
microorganisms.
If no carbon dioxide will be used for a long time, the CO
2
cylinder should be
closed to prevent the CO
2
regulator from malfunctioning.
The purity of used CO
2
must be 99.5%, or otherwise the sensitivity of CO
2
sensor will be reduced and the CO
2
filter will also be contaminated.
The inner door of the CO
2
Incubator cannot be condensated. Any observed