![RWD D180P Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/rwd/d180p/d180p_user-manual_1503442006.webp)
production, culture of hybridoma cells for antibody production, in vitro fertilization
(IVF), stem cells, tissue engineering, and drug screening.
1.4 Product features
Triple sterilization function. Continuous sterilization for 24 hours,
professional HEPA filtration system, and irregular sterilization;
Efficient, fast and easy high temperature sterilization cycle ensures not only
effective disinfection and sterilization, but also automatic running at night.
The entire sterilization cycle is less than 3 hours, and the total cycle is about
14 hours;
Temperature homogeneity. Thermoelements are located on the outer wall of
the incubator and is directly heated via six sides to ensure superior
temperature uniformity and reduce the time of heating sterilization.
The microprocessor control system is used to monitor the whole process of
high temperature sterilization in real time, and also display the status
information in different stages of high temperature sterilization cycle in real
time.
1.5 Equipment environmental requirements
The equipment operating environment is prepared according to the conditions
listed below to ensure the operability and safety of the system.
Detailed description
Working environment
Temperature: 5°C - 40°C
Humidity: 15%-95% (non-condensing)
Air pressure: 103.4 kPa
Storage environment
Temperature: -20°C - 60°C
Humidity: 10%-95% (non-condensing)
Air pressure: 50kPa-106kPa
Operating voltage
1)
210V-230V VAC 50/60HZ, 5A
2)
100V-120V VAC 50/60HZ, 10A
1.6 Specification
Parameters
Detailed description
Screen size
7-inch capacitive touch screen
External dimension:
Length × width × height
67cm*64.7cm*97cm
Internal dimension: Length 54.7cm*51cm*67.5cm
Содержание D180P
Страница 1: ...CO2 Incubator User Manual A...
Страница 19: ...Stacking rack Supporting feet...
Страница 20: ...3 2 Product show and ports introduction Product real shot display 1 2 3 4 5 6 7 8 9...