above alarm limit!
concentration alarm threshold
pressure valve of
cylinder is abnormal;
2
)
Adjust the preset
pressure of cylinder to
1.0-1.5bar.
Low CO
2
pressure!
CO
2
pressure is lower than 0.6bar;
1
)
Check whether the air
pressure valve of
cylinder is abnormal;
2
)
Replace the cylinder
and adjust the preset
pressure of cylinder to
1.0-1.5bar;
3
)
Check whether the air
supply line is abnormal.
High CO
2
pressure!
CO
2
pressure is higher than 1.8bar;
1)
Adjust the preset
pressure of cylinder to
1.0-1.5bar;
2)
If the pressure valve is
abnormal, close the air
source and replace the
pressure valve;
3)
If the above measures
are invalid, contact
after - sale
Relative
humidity
below alarm limit!
The current humidity is lower than
the lower limit of humidity alarm;
1)
Fill the humidity pan
with water.
Humidity sensor
error!
MCU have no communication with
humidity sensor
1)
Contact the after-sales
staff;
001 Voltage error,
please restart the
device or contact the
after-sales personnel!
Abnormal voltage
1)
Restart equipment;
2)
Contact the after-sales
staff;
002
CO
2
sensor
disconnected, please
restart the device or
contact the after-sales
personnel!
CO
2
sensor is not connected
1)
Restart equipment;
2)
Contact the after-sales
staff;
003 CO
2
sensor error,
please restart the
device or contact the
after-sales personnel!
CO
2
sensor is abnormal
1)
Restart equipment;
2)
Contact the after-sales
staff;
004 Temperature
sensor error, please
contact the after-sales
Temperature sensor is abnormal
1)
Contact the after-sales
staff;