background image

16

RADEX

  

AIRLINE FILTER

NOTE

Refer to your respirator instruction manual for pressure ranges for your respirators, increase 
the pressure according to the number of operators.

Check that ALL fittings and connections are tight.

 

YOUR FILTER IS NOW READY FOR USE.

INLET CONNECTION

The Radex

®

 Airline Filter has a 1” NPT inlet. Thread in a ball valve close to the inlet so 

the unit can be shut off for servicing. Connect inlet fittings and tighten. Once the Radex

®

 is 

completely assembled and connected to compressed air open supply line, check fittings for 
any air leaks and tighten accordingly.

DO NOT over tighten fittings, as over tightening could crack the 
filter body which may lead to an explosion and cause serious injury. 

Follow tightening directions from the maintenance section using a torque wrench.

BREATHING AIR SUPPLY HOSES AND FITTINGS

RPB

®

 air supply hoses and fittings must be used between the Radex

®

 (point of attachment) 

and the respirator breathing air connection at the wearer’s belt. The hose sections must be 
within the correct length and the amount of sections must be within the number specified in 
the breathing air pressure table in the instruction manual for the respirator being used. All 
connections should be sealed using liquid thread sealant. If the Radex

®

 Airline Filter is hard 

piped an isolation valve must be used to enable depressurization for servicing.

MAINTENANCE 
Drain Moisture Daily

Water will accumulate in the filter tank, this should be drained by opening the ball valve. 
This should be done each day. In very humid climates or if there is large amounts of water in 
the air supply it is recommended to leave the ball valve partially open to bleed the moisture. 
When fitted with auto drain assembly 04-924, moisture will automatically drain from the unit. 
Check it periodically as particles could cause blockages.

Filter cartridge replacement

The filter cartridge should be inspected weekly or more often depending on usage and the 
conditions of the air system in which the Radex

®

 Airline Filter is installed. The filter cartridge 

should be replaced after a period of 3 months based on a 40 hour week and typical usage. 
If the Pressure Gauge on the Super High Flow Regulator and the Pressure Gauge on the 
Manifold are reading more than 10psi apart, your cartridge is blocked and needs replacing.

The filter cartridge must be replaced immediately if the following exist:

1. 

The presence of odor and or tastes in the air being supplied to the respirator.

2. 

Presence of moisture at the outlet fittings.

3. 

Large pressure drop of 10psi or more across the filter.

WARNING

SETUP AND CARE 

CONTINUED

Содержание RADEX AIRLINE FILTER

Страница 1: ...d this manual and the flow control device Instruction Manual and follow the Product User Safety Instructions page 9 Manuals are regularly updated Make sure the most recent version of manual is availab...

Страница 2: ...Y STATEMENT 20 CONTENTS RPB Safety LLC is an ISO9001 certified company The following signal word and safety symbols are used in this manual and product labeling WARNING indicates a hazardous situation...

Страница 3: ...ual Property contained in this publication including copyright trademarks service marks trade secret and patent rights are reserved RPB Intellectual Property means any patent patented articles patent...

Страница 4: ...CTION PROVIDED AND LIMITATIONS Improper selection use or maintenance of respiratory protection products can result in injury life threatening delayed lung skin or eye disease or death This product is...

Страница 5: ...facturer instruction manual for recommendations on setting up the compressor for breathing air and for suitable in line air purifying sorbent beds and filters The compressor filters and oil level shou...

Страница 6: ...ure must be read from a reliable pressure gauge while the respirator has air flowing through it as required by applicable standards and regulations Refer to the instruction manual for the respirator f...

Страница 7: ...8 BREATHING TUBE ASSEMBLY AND FLOW CONTROL DEVICE 7 BREATHING AIR LINE 4 RADEX AIRLINE FILTRATION SYSTEM 5 GX4 GAS MONITOR 1 SUPPLIED AIR COMPRESSOR 2 SUPPLIED AIR LINE TO FILTER 6 POINT OF ATTACHMEN...

Страница 8: ...pecific airborne hazards with consideration of workplace and user factors and with an Assigned Protection Factor APF that meets or exceeds the required level for employee protection select welding fac...

Страница 9: ...tened securely Air will be released when pressure in the filter exceeds 150 psig 10 3 Bar If the hole is plugged with something other than the pressure relief valve excess pressure could cause the ves...

Страница 10: ...nant concentrations do not exceed those allowed by applicable regulations and standards for supplied air respirators Make sure the work area is ventilated and the work area and the breathing air suppl...

Страница 11: ...mount will only fit in two positions the arrows must align before locking Turn the filter in a clockwise direction and lock into position You will hear a click when the base is in the final position...

Страница 12: ...mounting onto the Radex Airline Filter Use thread sealant on all fittings Fit the complete regulator assembly onto the top of the Radex Airline Filter lid using a 17mm wrench Fit the pressure relief v...

Страница 13: ...Pressure Gauge into the 04 965 6 Outlet Manifold and tighten Thread all six Couplers into the 04 965 Manifold and tighten Thread the completed Six Outlet Manifold assembly into filter lid and tighten...

Страница 14: ...High Flow Regulator 04 962 are optional attachments for environments that are humid or have excessive moisture or have oil or other fine particulates in the air They can all be used together or indivi...

Страница 15: ...OR ASSEMBLY FIGURE 4 5 FIGURE 4 6 To fit the Auto Drain Unit screw the unit onto the elbow connector on the draining valve Make sure the unit is straight down so it will drain properly FIGURE 4 4 FITT...

Страница 16: ...nnections should be sealed using liquid thread sealant If the Radex Airline Filter is hard piped an isolation valve must be used to enable depressurization for servicing MAINTENANCE Drain Moisture Dai...

Страница 17: ...minants do not clean with volatile chemicals Irritation or sickness may result from lack of cleaning the inside of the Radex housing 5 Check the o rings 04 919 and replace if damaged 6 Insert a new fi...

Страница 18: ...18 RADEX AIRLINE FILTER 1 2 3 11 4 5 7 6 8 10 9 18 12 17 13 14 FIGURE 6 1 PARTS AND ACCESSORIES 16 19 22 28 23 27 26 29 30 15 20 21 31 32 24 25...

Страница 19: ...or 04 927 04 929 20 Brass Elbow 04 928 21 Auto Drain Unit 04 924 22 Auto Drain Assembly for 04 924 04 931 23 Super High Flow Regulator 04 962 24 Six Outlet Upgrade Kit inc 5 6 7 23 04 960 25 Six Outle...

Страница 20: ...ted defect occurs RPB will repair or replace the defective Product or a component of the Product in its sole discretion This repair or replacement remedy is the sole and exclusive remedy under this li...

Страница 21: ...21 NOTES...

Страница 22: ...22 RADEX AIRLINE FILTER NOTES...

Страница 23: ...23 NOTES...

Страница 24: ...e value AIR QUALITY MONITORING Do you need an intelligent gas monitor that can give you complete confidence in the air you and your employees are breathing The RPB GX4 Gas Monitor has the ability to d...

Отзывы: