Rosemount Tank Radar Rex
Chapter 4 Service
4-16
Service Manual
308012En, Ed.2
May 2013
TIC/TMC Configuration
The TMC board is configured depending on what type of sensor that is
used. The jumpers are installed at factory and do normally not need to be
changed. If you need to change TMC board configuration, for example if
other sensor types than originally specified will be used, do the following:
Note!
Make sure that power is turned off.
1
Check if there is any screw
that is sealed. Contact Rose-
mount Tank Gauging before
breaking the seal if warranty is
still valid. Completely remove
the seal so that it does not
damage the threads. Loosen
the screws and remove the
cover.
2
Remove the PCB locking from
the transmitter head and
remove the TIC/TMC board.
3
Carefully separate the TMC
from the TIC board.
PCB locking
TMC
TIC
TMC
TIC
Содержание TankRadar Rex DAU 2100
Страница 2: ......
Страница 3: ...www rosemount tg com Service manual Second edition Copyright May 2013 Rosemount Tank Radar AB ...
Страница 8: ...Rosemount TankRadar Rex Contents TOC 4 Service Manual 308012En Ed 2 May 2013 ...
Страница 12: ...Rosemount Tank Radar Rex Chapter 1 System Description 1 4 Service Manual 308012En Ed 2 May 2013 ...
Страница 44: ...Rosemount Tank Radar Rex Chapter 3 Device Description 3 30 Service Manual 308012En Ed 2 May 2013 ...
Страница 116: ...Rosemount Tank Radar Rex Chapter 4 Service 4 72 Service Manual 308012En Ed 2 May 2013 ...
Страница 124: ...Rosemount Tank Radar Rex Chapter 5 Calibration 5 8 Service Manual 308012En Ed 2 May 2013 ...
Страница 132: ...Rosemount Tank Radar Rex Chapter 6 Troubleshooting 6 8 Service Manual 308012En Ed 2 May 2013 ...
Страница 134: ...Rosemount Tank Radar Rex Chapter 7 Spare Parts 7 2 Service Manual 308012En Ed 2 May 2013 ...
Страница 146: ...Rosemount Tank Radar Rex Chapter 9 Technical Data 9 10 Service Manual 308012En Ed 2 May 2013 ...
Страница 152: ...Rosemount Tank Radar Rex Index 1 6 Service Manual 308012En Ed 2 May 2013 ...
Страница 153: ......