5-7
Rosemount Tank Radar Rex
Chapter 5 Calibration
Service Manual
308012En, Ed.2
May 2013
6
Click the
Write new calibration data to RTG
button in order to save the
current calibration data.
Note!
By clicking the Write new calibration data to RTG button, the Rosemount
Level values in the Calibration Data window are recalculated and the old
Calibration Data is replaced.
Now you can check the calibration result by opening the
Calibration
window again:
Note that all measured values are adjusted according to the calculated
Calibration Distance. In the
Calibration Data
window you can also see
that the level values measured by the transmitter are adjusted. Of
course, the hand dipped levels are unaltered.
Содержание TankRadar Rex DAU 2100
Страница 2: ......
Страница 3: ...www rosemount tg com Service manual Second edition Copyright May 2013 Rosemount Tank Radar AB ...
Страница 8: ...Rosemount TankRadar Rex Contents TOC 4 Service Manual 308012En Ed 2 May 2013 ...
Страница 12: ...Rosemount Tank Radar Rex Chapter 1 System Description 1 4 Service Manual 308012En Ed 2 May 2013 ...
Страница 44: ...Rosemount Tank Radar Rex Chapter 3 Device Description 3 30 Service Manual 308012En Ed 2 May 2013 ...
Страница 116: ...Rosemount Tank Radar Rex Chapter 4 Service 4 72 Service Manual 308012En Ed 2 May 2013 ...
Страница 124: ...Rosemount Tank Radar Rex Chapter 5 Calibration 5 8 Service Manual 308012En Ed 2 May 2013 ...
Страница 132: ...Rosemount Tank Radar Rex Chapter 6 Troubleshooting 6 8 Service Manual 308012En Ed 2 May 2013 ...
Страница 134: ...Rosemount Tank Radar Rex Chapter 7 Spare Parts 7 2 Service Manual 308012En Ed 2 May 2013 ...
Страница 146: ...Rosemount Tank Radar Rex Chapter 9 Technical Data 9 10 Service Manual 308012En Ed 2 May 2013 ...
Страница 152: ...Rosemount Tank Radar Rex Index 1 6 Service Manual 308012En Ed 2 May 2013 ...
Страница 153: ......