M
ODEL
755
O
XYGEN
A
NALYZER
2-4
748183-H
Rosemount Analytical
November 1997
U5
R2 R1
R3
R4
R5
R6
R8
R9
CR2
U6
I G O
U3
U2
1
2
3
4
I
G
O
C5
C4
C2
U4
I G O
C3 CR1 C1
U1
J1
Current Output Board
RECORDER OUTPUT
Selector Plug
R2
R1
R67
R68
R78
R73
R64
R63
5V
1V
100mV
10mV
P
OTENTIOMETRIC
O
UTPUT
1. Insert RECORDER OUTPUT Selector Plug (Figure 2-2) in position appropriate
to the desired output: 10 mV, 100 mV, 1 V, or 5 V.
2. On TB2, Figure 2-1, connect leads of shielded recorder cable to MV+ and
COM terminals.
3. Connect free end of output cable to appropriate terminals of recorder or other
potentiometric device:
a. For device with a span of 0 to 10mV, 0 to 100mV, 0 to 1V, or 0 to 5V, connect
cable directly to input terminals of the device, making sure polarity is correct.
b. For device with intermediate span, i.e., between the specified values, connect
cable to device via a suitable external voltage divider, as shown in Figure 2-3.
F
IGURE
2-2. C
ONTROL
B
OARD
- A
DJUSTMENT
L
OCATIONS
Содержание 755
Страница 1: ...Rosemount Analytical MODEL 755 OXYGEN ANALYZER INSTRUCTION MANUAL 748183 H ...
Страница 16: ...MODEL 755 OXYGEN ANALYZER Rosemount Analytical 748183 H November 1997 P 8 NOTES ...
Страница 22: ...MODEL 755 OXYGEN ANALYZER 1 6 748183 H Rosemount Analytical November 1997 NOTES ...
Страница 40: ...MODEL 755 OXYGEN ANALYZER 2 18 748183 H Rosemount Analytical November 1997 NOTES ...
Страница 50: ...MODEL 755 OXYGEN ANALYZER 3 10 748183 H Rosemount Analytical November 1997 NOTES ...
Страница 62: ...MODEL 755 OXYGEN ANALYZER 5 10 748183 H Rosemount Analytical December 1997 NOTES ...
Страница 90: ... blank ...
Страница 92: ... blank ...
Страница 94: ... blank ...