M
ODEL
755
O
XYGEN
A
NALYZER
2-6
748183-H
Rosemount Analytical
November 1997
755
Analyzer
Recorder
Controller
Remote
Indicator
+
-
+
-
+
-
+
-
mA
F
IGURE
2-4. M
ODEL
755 C
ONNECTED
T
O
D
RIVE
C
URRENT
O
UTPUT
-A
CTIVATED
O
UTPUT
D
EVICES
2.4.3 O
UTPUT
C
ONNECTIONS
, I
NITIAL
S
ETUP FOR
D
UAL
A
LARM
O
PTION
If so ordered the analyzer is factory-equipped with alarm output. Alternatively the
alarm feature is obtainable by subsequent installation of the Alarm Kit.
A
LARM
O
UTPUT
C
ONNECTIONS
The alarm output provides two sets of relay contacts for actuation of alarm or process-
control functions. Leads from the customer-supplied external alarm system connect to
terminals on the Alarm Assembly, see Figure 2-1.
Note the following recommendations:
1. A fuse should be inserted into the line between the customer-supplied power supply and the alarm
relay terminals on the Alarm Relay Assembly.
2. If the alarm contacts are connected to any device that produces radio frequency interference (RFI),
it should be arc-suppressed. The 858728 Arc Suppressor is recommended.
3. If at all possible, the analyzer should operate on a different AC power source, to avoid RFI.
A
LARM
R
ELAY
C
HARACTERISTICS
The Alarm 1 and Alarm 2 outputs of the 638245 Alarm Relay Assembly are provided
by two identical single-pole double-throw relays. Relay contacts are rated at:
5 amperes, 240 VAC resistive
1 ampere, 240 VAC inductive
5 amperes, 120 VAC resistive
3 amperes, 120 VAC inductive
5 amperes, 30 VDC resistive
3 amperes, 30 VDC inductive
Содержание 755
Страница 1: ...Rosemount Analytical MODEL 755 OXYGEN ANALYZER INSTRUCTION MANUAL 748183 H ...
Страница 16: ...MODEL 755 OXYGEN ANALYZER Rosemount Analytical 748183 H November 1997 P 8 NOTES ...
Страница 22: ...MODEL 755 OXYGEN ANALYZER 1 6 748183 H Rosemount Analytical November 1997 NOTES ...
Страница 40: ...MODEL 755 OXYGEN ANALYZER 2 18 748183 H Rosemount Analytical November 1997 NOTES ...
Страница 50: ...MODEL 755 OXYGEN ANALYZER 3 10 748183 H Rosemount Analytical November 1997 NOTES ...
Страница 62: ...MODEL 755 OXYGEN ANALYZER 5 10 748183 H Rosemount Analytical December 1997 NOTES ...
Страница 90: ... blank ...
Страница 92: ... blank ...
Страница 94: ... blank ...