I
NSTALLATION
748183-H
2-9
Rosemount Analytical
November 1997
band. Both potentiometers must remain at this setting throughout calibration of
the alarm setpoint adjustments.
4. Connect an ohmmeter to relay terminals on 638254 Alarm Relay Assembly to
verify when alarms have been energized.
5. Calibration of ALARM 1, HIGH.
a. Rotate setpoint adjustment, R64, fully counter-clockwise.
b. Adjust SPAN control to give a display or recorder reading exactly fullscale. If
the fullscale setting cannot be reached, set to a reading higher than the
desired alarm setpoint.
c. Set ALARM 1 calibration adjustment, R63, to its clockwise limit (Figure 2-2).
Rotate R63 counter-clockwise the minimum amount required to energize
ALARM 1, relay K1. Verify that the alarm has been energized with the
ohmmeter on the relay contacts (Figure 2-7).
F
IGURE
2-7. A
LARM
R
ELAY
O
PTION
S
CHEMATIC
D
IAGRAM
6. Calibration of ALARM 2, LOW.
a. Rotate setpoint adjustment, R68, fully counter-clockwise.
2. RELAYS SHOWN IN ENERGIZED POSITION.
1. CR1 AND CR2 ARE ANY 600 V, 1 AMP DIODE.
NOTES:
+15V
-15V
ALARM 1
COMMON
ALARM 2
COMMON
J5
1
2
4
6
14
CR1
13
K1
1
5
9
12
8
NO
COM
NC
RESET
ALARM 1
CR2
13
K2
1
5
9
12
8
NO
COM
NC
RESET
ALARM 2
14
Содержание 755
Страница 1: ...Rosemount Analytical MODEL 755 OXYGEN ANALYZER INSTRUCTION MANUAL 748183 H ...
Страница 16: ...MODEL 755 OXYGEN ANALYZER Rosemount Analytical 748183 H November 1997 P 8 NOTES ...
Страница 22: ...MODEL 755 OXYGEN ANALYZER 1 6 748183 H Rosemount Analytical November 1997 NOTES ...
Страница 40: ...MODEL 755 OXYGEN ANALYZER 2 18 748183 H Rosemount Analytical November 1997 NOTES ...
Страница 50: ...MODEL 755 OXYGEN ANALYZER 3 10 748183 H Rosemount Analytical November 1997 NOTES ...
Страница 62: ...MODEL 755 OXYGEN ANALYZER 5 10 748183 H Rosemount Analytical December 1997 NOTES ...
Страница 90: ... blank ...
Страница 92: ... blank ...
Страница 94: ... blank ...