C
IRCUIT
A
NALYSIS
748183-H
6-3
Rosemount Analytical
December 1997
-15V
R74
590K
-
+3
R75
210K
R76
37.4K
C40
2200uF
C37
1.0uF
Q6
R80
10K
R79
10K
R81
56.2
C38
.18uF
R87
10K
R78
249K
R77
10K
TO
COMPARATOR
R83
63.4K
R84
169K
R82
9.09K
RT1
-15V
-15V to 1.88V
±
0.3V
+15V
+15V
-15V
INPUT FROM
MULTIVIBRATOR
-15V
OFF
OFF
OFF
-2.3V
+2.3V
6 Hz
T2
U1-C
CASE BOARD
120 V
RMS
T1
19 VAC
TO POWER
SUPPLY
19 VAC
12
R67
10K
R72
4.75K
C38
.18uF
CR10
CR9
-15V
R71
21.5K
R69
2 M
R68
3.3K
R70
20M
R73
20M
R74
590K
+15V
-
+
U1-A
-
+
U1-B
-
+
U1-C
-
+
U1-D
R75
210K
R85
11.0K
R86
20M
R76
37.4K
C40
2200uF
C37
1.0uF
C39
.01uF
R83
63.4K
R84
169K
R82
9.07K
RT1
-15V
R78
249K
Q6
R80
10K
R79
10K
R81
56.2
C38
.18uF
R87
10K
R77
10K
CR11
T2
+15V
CASE BOARD
generator, probably -188
±
0.03 V.
The on-off effect of the comparators to the OR circuit results in application of a
positive-going pulse (from -15 V to -1.89 V) to the temperature bridge at the rate of
120 pulses per second.
Capacitor C36 is added to the input circuit to delay the incoming AC signal so that the
pulses will occur at or just after the line frequency crossover point.
Circuits for a ramp generator and a temperature-sensing bridge are part of the case
heater control circuit of Figures 6-2 and 6-3.
F
IGURE
6-2. C
ASE
H
EATER
C
ONTROL
C
IRCUIT
F
IGURE
6-3. R
AMP
G
ENERATOR
Содержание 755
Страница 1: ...Rosemount Analytical MODEL 755 OXYGEN ANALYZER INSTRUCTION MANUAL 748183 H ...
Страница 16: ...MODEL 755 OXYGEN ANALYZER Rosemount Analytical 748183 H November 1997 P 8 NOTES ...
Страница 22: ...MODEL 755 OXYGEN ANALYZER 1 6 748183 H Rosemount Analytical November 1997 NOTES ...
Страница 40: ...MODEL 755 OXYGEN ANALYZER 2 18 748183 H Rosemount Analytical November 1997 NOTES ...
Страница 50: ...MODEL 755 OXYGEN ANALYZER 3 10 748183 H Rosemount Analytical November 1997 NOTES ...
Страница 62: ...MODEL 755 OXYGEN ANALYZER 5 10 748183 H Rosemount Analytical December 1997 NOTES ...
Страница 90: ... blank ...
Страница 92: ... blank ...
Страница 94: ... blank ...