Reference Manual
00809-0100-4030, Rev DA
June 2007
Rosemount 2120
B-12
ATEX Intrinsically Safe Approval
I1
Certificate:
Sira
05ATEX2130X
Intrinsic Safety:
ATEX Marking
II 1 G D
EEx ia IIC T5 (T
amb
-40°C to +80°C)
Ui=15 V, Ii=32 mA, Pi=0.1 W, Ci=12 nF, Li=0.06 mH
NOTE
A NAMUR isolating amplifier must be used for intrinsic safety.
Instructions specific to hazardous area installations
Model numbers covered: 2120XXXCXI1XX
(“X” indicates options in construction, function and materials.)
The following instructions apply to equipment covered by certificate number
Sira 05ATEX2130X:
1. The 2120 I.S. may be used in a hazardous area with flammable gases and vapors with
apparatus groups IIC, IIB and IIA and with temperature classes T1, T2, T3, T4 and T5.
2. The apparatus electronics is only certified for use in ambient temperatures in the range
of -40ºC to +80ºC. It should not be used outside this range. However, the switch may be
located in the process medium which may be at a higher temperature than the
electronics but must not be higher than the Temperature Class for the respective process
gas/medium.
It is a special condition of the certification that the temperature of the electronics housing
is in the range of -40ºC to +80ºC. It must not be used outside this range. It will be
necessary to limit the external ambient temperature if the process temperature is high.
(See Technical Data below.)
3. Suitably trained personnel shall carry out installation in accordance with the applicable
code of practice.
4. The user should not repair this equipment.
5. If the equipment is likely to come into contact with aggressive substances, it is the
responsibility of the user to take suitable precautions that prevent it from being adversely
affected, thus ensuring that the type of protection is not compromised.
Aggressive Substances
– e.g. acidic liquids or gases that may attack metals or
solvents that may affect polymeric materials.
Suitable Precautions
– e.g. regular checks as part of routine inspections or establishing
from the material’s data sheet that it is resistant to specific chemicals.
Process Temperature
Temperature classes
80°C
T1, T2, T3, T4, T5
115°C
T1, T2, T3, T4,
150°C
T1, T2, T3
Содержание 2120
Страница 2: ......
Страница 4: ...Reference Manual 00809 0100 4030 Rev DA June 2007 Rosemount 2120 ...
Страница 8: ...Reference Manual 00809 0100 4030 Rev DA June 2007 Rosemount 2120 TOC 4 ...
Страница 68: ...Reference Manual 00809 0100 4030 Rev DA June 2007 Rosemount 2120 B 20 ...
Страница 69: ......
Страница 70: ......
Страница 71: ......