Accura 2350-DCM User Guide
Chapter 2 Accura 2350-DCM Operation/Setup
ⓒ
2018 Rootech Inc. All Rights Reserved
Page 15
Measurement and Event
Specifications
Item
Description
Voltage Measurement
Sampling time
1 ms
Aggregation of
measurement data
Aggregates data every 20 ms
1
based
on the data measured over 1 ms
Window aggregation:1
–
32 ms
(default: 20 ms)
Aggregates data every 0.5 sec
2
based on the data measured over 20 ms
Measurement accuracy
±
0.5% RD +
±
0.03% FS
Accura 2350-DCM-50V
FS=50V
Accura 2350-DCM-500V FS=500V
Measuring range
(accuracy guaranteed)
3
DC 3
–
50 V
Accura 2350-DCM-50V
DC 30
–
500 V
Accura 2350-DCM-500V
Minimum measured value
3,
4
DC 0.2 V
Accura 2350-DCM-50V
configurable
DC 2 V
Accura 2350-DCM-500V
Maximum measured value
DC 55 V
Accura 2350-DCM-50V
DC 550 V
Accura 2350-DCM-500V
Current Measurement
Sampling time
1 ms
Aggregation of
measurement data
Aggregates data every 20 ms
1
based
on the data measured over 1 ms
Window aggregation:1
–
32 ms
(default: 20 ms)
Aggregates data every 0.5 sec
2
based on the data measured over 20 ms
Measurement accuracy
±
0.5% RD +
±
0.06% FS
FS=80A
Measuring range
[accuracy guaranteed]
DC 0.8
–
80 A
Minimum measured value
4
DC 0.08 A
Configurable
Maximum measured value
DC 100 A
Temperature drift
5
±
0.02% FS/
℃
typ.
Calibration temperature: 25
℃
Power / Energy Measurement
Aggregation of
measurement data
Power/Energy measurement data aggregated for 0.5 sec
2
Measurement accuracy
±
1.0% RD +
±
0.15% FS
Based on voltage/current measurement
accuracy
Event
Number of event channels
5 (channel 1,2,3,4,5)
Event type
Off / Over
6
/ Under
7
selectable
(default: Off)
Configurable for each channel
Monitored signal
Voltage / Current selectable
Configurable for each channel
Operation linked with
Digital Output
DO operation mode
Disable / Enable
Configurable for each channel
Содержание ACCURA 2350-DCM
Страница 1: ......