10-2. ERROR CODE LIST
E0100
[ Main heater thermocouple disconnection ]
- Remove cables from temperature controller
And check disconnection or defective on
wires.
Replace if defective.
- Check temperature controller.
E0101
[ Main heater trouble ]
- Check main heater resistance under no
power source supplied.
- Check SSR (solid state relay for main heater)
at maintenance window by tapping SSR
button to check if LED activated or not.
Replace SSR if not.
E0200
[ Afterburner thermocouple disconnection ]
- Remove cables from control board and
check disconnection or defective on wires.
Replace if defective.
E0400
[ Afterburner trouble ]
- Check heater resistance under no power
source supplied.
- Check SSR (solid state relay for afterburner
heater) at maintenance window by tapping
SSR button to check if LED activated or not.
Replace SSR if not.