Troubleshooting
R&S
®
TSME
63
User Manual 1514.6542.02 ─ 10
Currently the R&S
TSME does not answer to ICMP echo requests (PING). How-
ever, for ViCom users, a R&S
TSME
Ping
tool is available after installation of R&S
ViCom (
C:\RuS\ViCom_15.0\tools\TSME\TsmePing.exe
).
●
Is the PC configured to use the subnet
192.168.0.0
with a different IP address to
the R&S
TSME (default:
192.168.0.2
)? (See also
the LAN Interface on the Host PC"
●
Are
other applications
trying to access the R&S
TSME at the same time (e.g.
R&S
ROMES, R&S
NESTOR, R&S
ViCom, R&S
TSME Device Manager)? Check
the Windows TaskManager and look for associated processes. If necessary, stop
the processes.
●
Is the host using
other network connections
at the same time, such as VPN?
R&S
TSME does not work when the PC is connected to VPN.
●
Is the
connecting cable
between the host and the R&S
TSME the one provided
with the R&S
TSME? Make sure you are using a quality CAT5e or better Gigabit
Ethernet conform cable. RJ45 connectors must be shielded, not plastic.
●
After switching the R&S
TSME off and back on, is the
[power] LED
on the
R&S
TSME
blinking
while the
[state] LED
is
off
? This indicates that the TSME
cannot boot. Contact Rohde
&
Schwarz Support Center.
6.3
Solving Other Miscellaneous Problems
In case measurement problems occur with the R&S
TSME, check the "Device Analysis
Output" table in the R&S
TSME Device Manager for any errors that may have been
detected as a first measure (see
Problem: R&S ROMES does not detect the R&S TSME
.............................................. 63
Problem: The R&S TSME Device Manager does not detect the R&S TSME
Problem: The options in the R&S TSME Device Manager are lost
...............................64
Problem: MIMO operation on multiple R&S TSMEs fails
.............................................. 64
Problem: Error message: "No calibration data"
.............................................................64
Problem: R&S TSME does not provide measurement data
..........................................64
Problem: The [STATE] LED is blinking red
....................................................................64
Problem: Instrument shuts down at lower temperature than expected
Problem: R&S ROMES does not detect the R&S TSME
Possible solutions:
●
Restart R&S ROMES, while the R&S
TSME is connected and switched on.
●
Restart the PC and then start R&S ROMES, while the R&S
TSME is connected
and switched on.
If you switch the instruments off and on again, be sure to switch on the R&S
TSME
before you start R&S ROMES!
Problem: The R&S TSME Device Manager does not detect the R&S TSME
Solution:
Solving Other Miscellaneous Problems
Содержание TSME
Страница 1: ...R S TSME Ultra Compact Drive Test Scanner User Manual User Manual Version 10 1514654202 Z2...
Страница 6: ...Contents R S TSME 4 User Manual 1514 6542 02 10 Annex 65 A Available Cellular Bands 65 Index 72...
Страница 59: ...Maintenance R S TSME 57 User Manual 1514 6542 02 10...
Страница 75: ...Index R S TSME 73 User Manual 1514 6542 02 10 Troubleshooting LAN connection 60 U Updates Firmware 56...