Troubleshooting
R&S
®
TSME
59
User Manual 1514.6542.02 ─ 10
Connected mode
As soon as the host PC application (R&S
ROMES or R&S
NESTOR) or customer
application based on the ViCom interface accesses the R&S
TSME via the LAN inter-
face, the R&S
TSME switches to
connected mode
.
After establishing the connection to the R&S
TSME, the first thing the host PC applica-
tion does is to send the required application firmware to the R&S
TSME. The
R&S
TSME is then ready for measurement commands sent from R&S
ROMES,
R&S
NESTOR, or R&S
ViCom.
The [STATE] LED is green.
Measuring mode
The instrument switches to
measuring mode
immediately after receiving any measur-
ing command from the host PC software.
In measuring mode the [STATE] LED blinks green (rapidly).
Standby mode
When you press the "On/Off" key on the rear panel of the R&S
TSME, the instrument
switches to
standby mode
. Program execution on the instrument is stopped immedi-
ately, but the instrument is still under power connection. The [Power] LED is yellow, all
other LEDs are off.
If you remove the power supply in standby mode and reconnect it later, the R&S
TSME
switches back to standby mode and does not automatically boot. You must press the
"On/Off" key first to start booting.
Error mode
If an error occurs during operation and no measurements are possible, the R&S
TSME
falls into
error mode
.
The [STATE] LED turns red. If possible, warnings or errors are indicated in the "Device
Analysis Output" table of the R&S
TSME Device Manager (see
The following diagram shows the order in which the R&S
TSME goes through the vari-
ous instrument modes.
Overview of Instrument Modes
Содержание TSME
Страница 1: ...R S TSME Ultra Compact Drive Test Scanner User Manual User Manual Version 10 1514654202 Z2...
Страница 6: ...Contents R S TSME 4 User Manual 1514 6542 02 10 Annex 65 A Available Cellular Bands 65 Index 72...
Страница 59: ...Maintenance R S TSME 57 User Manual 1514 6542 02 10...
Страница 75: ...Index R S TSME 73 User Manual 1514 6542 02 10 Troubleshooting LAN connection 60 U Updates Firmware 56...