Remote Control Commands
R&S
®
RTM3000
779
User Manual 1335.9090.02 ─ 08
WGENerator:MODulation:FSK:HFREquency
<HoppingFrequency>
Sets the second frequency of the FSK-modulated signal.
Parameters:
<HoppingFrequency> The range depends on the signal type:
Sinus: 100mHz - 25MHz
Sinc/Triangle/Ramp/Exponential: 100mHz - 1MHz
Rectangle/Pulse/Arbitrary: 100mHz - 10MHz
WGENerator:MODulation:FSK:RATE
<Rate>
Sets the rate at which signal switches between the carrier frequency and the hopping
frequency.
Parameters:
<Rate>
Range:
0.1 to 1E6
Default unit: Hz
WGENerator:MODulation:RAMP:POLarity
<Polarity>
Sets the polarity for the ramp function for a modulation waveform.
Parameters:
<Polarity>
POSitive | NEGative
16.14.1.5
Sweep Settings
............................................................................................ 779
........................................................................................... 779
............................................................................................. 780
............................................................................................ 780
........................................................................................780
WGENerator:SWEep:FEND
<StopFrequency>
Sets the stop frequency of the sweep signal.
Parameters:
<StopFrequency>
WGENerator:SWEep:FSTart
<StartFrequency>
Sets the start frequency of the sweep signal.
Parameters:
<StartFrequency>
Signal Generation (Option R&S
RTM-B6)
www.allice.de