Waveform Analysis
R&S
®
RTM3000
134
User Manual 1335.9090.02 ─ 08
"Equal"
Finds rise/fall times equal to the reference "Rise/Fall Time" if "Varia-
tion" Δt = 0.
If "Variation" ≠ 0, the setting finds rise/fall times within the range time
± Δt.
"Not equal"
Finds rise/fall times unequal to the reference value if "Variation" Δt =
0.
If "Variation" ≠ 0, the setting finds rise/fall times outside the range
time ± Δt.
Remote command:
Rise/Fall Time
Sets the reference rise or fall time, the nominal value for comparisons.
Remote command:
Variation
Sets a range Δt to the reference "Rise/Fall Time" if comparison is set to "Equal" or "Not
equal". The instrument finds rise/fall times inside or outside the range width ± Δt.
Remote command:
6.5.7
Runt Setup
The runt search finds pulses lower than normal in amplitude. The amplitude crosses
the first threshold twice without crossing the second one. In addition to the threshold
amplitudes, you can define a time limit for the runt in the same way as for width search:
runts with exact width, shorter or longer than a given time, or runts inside or outside the
allowable time range.
► [Search] > "Search Type" = "Runt" > "Setup"
Search
www.allice.de