Applications
R&S
®
RTM3000
206
User Manual 1335.9090.02 ─ 08
4. Set the amplitude for the measurement.
5. If necessary, open the "Setup" dialog to refine the settings:
a) Set an "Amplitude Profile" for systems with sensitive circuits.
b) Set the "Maximum Phase" of the system.
c) Set a "Meas. Delay" for systems that need more time to adapt to a new fre-
quency.
6. Press "Run" to start the measurement.
7. If necessary, press "Save" to save the measurement results to a file.
8.8.3
Bode Plot Window Controls
The Bode plot window provides the most important function to set up the plot and to
run the test.
Start
Sets the start frequency of the sweep for the Bode plot.
Remote command:
Stop
Sets the stop frequency of the sweep for the Bode plot.
Remote command:
Points
Selects the number of points per decade that are measured from a predefined list. You
can also set a different value from the predefined ones in the "Setup" menu.
Gen
Enables the in-build generator. This allows you to start a frequency sweep for a defined
frequency range.
Ampl.
Sets a fixed amplitude.
Input
Selects the channel for the input signal of the DUT.
Remote command:
Bode Plot (Option R&S
RTM-K36)