Serial Bus Analysis
R&S
®
RTM3000
329
User Manual 1335.9090.02 ─ 08
Threshold High
Sets the upper threshold.
Remote command:
Threshold Low
Sets the lower threshold.
Remote command:
Bit Rate
Sets the bit rate mode to high speed (100 kBit/s), low speed (12.5 kBit/s) or a user
defined value.
Remote command:
Decode Format
Sets the decoding data format for the ARINC 429 bus: Data, SSM+Data, SSM+Data
+SDI, Data+SDI.
Remote command:
12.9.3
ARINC 429 Trigger
Before you set up the trigger, make sure that the bus is configured correctly. See
Chapter 12.9.2, "ARINC 429 Configuration"
To trigger on ARINC 429 signals:
1. Press the [Protocol] key in the Analysis area of the front panel.
2. Select the bus that is configured for ARINC 429.
3. Select "Trigger".
This selection has several effects:
● Enables decoding, if necessary.
● Sets the "Trigger Type" to "Serial Bus" and the trigger source to the selected
bus.
● Displays the "ARINC
Trigger" condition in the dialog box, below the protocol
setup.
4. At "ARINC
Trigger", select the required trigger type:
● "Word": word start or word stop
● "Error": combination of protocol errors
● "Label": on a spcified label.
● "Label and Data": combination of data and label condition
ARINC 429 (Option R&S
RTM-K7)