CALCulate:MATH<m>:TRACk:DPULse[:ENABle]
<DoublePulseEnable>
Available for bipolar signals. Considers the effect of the double pulse, which arises due
to the inversion.
Suffix:
<m>
.
1..5
Parameters:
<DoublePulseEnable>ON | OFF
CALCulate:MATH<m>:TRACk:THReshold:HYSTeresis
<Hysteresis>
Sets the hysteresis for correct edge detection.
Suffix:
<m>
.
1..5
Parameters:
<Hysteresis>
Default unit: V
CALCulate:MATH<m>:TRACk:THReshold[:UPPer]
<ThresholdLevel>
Sets the threshold for unipolar track sources, or the upper level for bipolar sources.
The signal is measured at the crossing points of the signal with the threshold.
Suffix:
<m>
.
1..5
Parameters:
<ThresholdLevel>
Default unit: V
CALCulate:MATH<m>:TRACk:THReshold:LOWer
<ThresholdLevel>
Sets the lower level for bipolar track sources, which determines the crossing points on
negative pulses.
Suffix:
<m>
.
1..5
Parameters:
<ThresholdLevel>
Default unit: V
16.6.3
Reference Waveforms
In all REFCurve:... commands, the suffix <m> selects the reference waveform.
To get the waveform data, use the following commands:
●
●
●
Waveform Analysis