Operating Concepts
R&S
®
NRP-Zxx
31
Getting Started 1176.8822.02 ─ 03
Setup
Figure 6-3: Setup with an R&S NRP2 base unit
1 = Signal source
2 = R&S NRP
‑
Z power sensor
3 = SENSOR connector of the R&S NRP2
4 = R&S NRP2 base unit
Incorrectly connecting/disconnecting the R&S NRP
‑
Z power sensors may damage the
power sensors or lead to erroneous results.
Ensure that you connect/disconnect your power sensors as described in
Starting a measurement
1. Connect the power sensor to the SENSOR connector of the R&S NRP2 as shown
2. Preset the R&S NRP2.
3. Execute zeroing.
Note:
Turn off all test signals before zeroing. An active test signal during zeroing
causes an error.
4. Connect the power sensor to the signal source.
The result window indicates the result (in dBm) obtained with sensor A.
5. If necessary, perform further settings.
R&S NRP2