Preparing for Use
R&S
®
FSVA/FSV
59
Quick Start Guide 1321.3066.02 ─ 05
Deactivating the automatic login function
To deactivate the automatic login function, perform the following steps:
1.
Select the "Windows" icon in the toolbar to access the operating system of the
R&S
Chapter 3.5.4, "Accessing the Start Menu"
2. In the "Start" menu, select "Run".
The "Run" dialog box is displayed.
3. Enter the command
C:\R_S\INSTR\USER\NO_AUTOLOGIN.REG
.
4. Press the ENTER key to confirm.
The automatic login function is deactivated. The next time you switch on the
instrument, you are prompted to enter your user name and password before
the firmware is started.
Adapting the automatic login function to a new password
If you change the "Instrument" user's (administrator's) password, which is used
during automatic login, this function no longer works. You must adapt the settings
for the command that activates the auto login function first.
1. Open the
C:\R_S\INSTR\USER\NO_AUTOLOGIN.REG
file in any text editor
(e.g. Notepad).
2. In the line
"DefaultPassword"="894129"
, replace the default password
(
894129
) by the new password which is to be used for automatic login.
3. Save the changes to the file.
Re-activating the automatic login function
1. In the "Start" menu, select "Run".
The "Run" dialog box is displayed.
2. Enter the command
C:\R_S\INSTR\USER\AUTOLOGIN.REG
.
3. Press the ENTER key to confirm.
The automatic login function is re-activated. It will be applied the next time the
instrument is rebooted.