Basic Measurement Examples
R&S
®
FSVA/FSV
137
Quick Start Guide 1321.3066.02 ─ 05
frequency). The resolution bandwidth and the frequency offset must be selected
in such a manner that the instantaneous frequency is located in the linear part of
the filter edge. As a result, the frequency variation of the FM-modulated signal is
transformed into an amplitude variation that can be displayed on screen in zero
span.
Displaying the AF of an FM-Modulated Carrier
Test setup
Table 6-8: Signal generator settings (e.g. R&S SMU)
Frequency
128 MHz
Level
-20 dBm
Modulation
FM 0 kHz deviation (i.e. FM modulation is
deactivated), 1 kHz AF
1. Set the signal analyzer to the default state by pressing the PRESET key.
The R&S
FSVA/FSV is set to its default state.
2. Set the center frequency to 127.50 MHz and the span to 300 kHz.
a) Press the FREQ key and enter
127.50 MHz
.
b) Press the SPAN key and enter
300 kHz
.
3. Set the resolution bandwidth to 300 kHz.
a) Press the BW key.
b) Press the "Res BW Manual" softkey and enter
300 kHz
.
c) Press the "Video BW Manual" softkey and enter
30 kHz
.
4. Set the display range to 20 dB and shift the filter trace to the center of the
screen.
a) Press the AMPT key.
b) Press the "Range" softkey
c) Press the "Range Log Manual" softkey and enter
20 dB
.
d) Press the "Up
↑
" softkey.
e) Press the "More" softkey.
f) Switch the "Grid" softkey to "Rel".
g) Press the "Up
↑
" softkey.
h) Press the "Ref Level" softkey.
Measurements in Zero Span