3.1.6
Using an Optional DC Power Supply
When only DC power is available, for example from a battery or in a vehicle, an
optional DC power supply adapter (R&S
FSV-B30) can be connected to the
R&S
FSVA/FSV to operate the instrument with a DC voltage of 10
V to 28
V.
For installation instructions see the option's installation guide.
Shock hazard due to power supply
If you use external power supplies to supply the equipment with DC safety
extra low voltage (SELV), make sure the requirements for reinforced or
double insulation are met, according to DIN/EN/IEC 61010 (UL61010, CAN
CSA C22.2 No. 61010) or DIN/EN/IEC 60950 (UL60950, CAN CSA C22.2
No. 60950). Otherwise you may be injured due to electrical shock.
Before switching on the R&S
FSVA/FSV, switch on the DC power supply:
Switching the DC power supply on
1. Connect the R&S
FSVA/FSV DC power supply device to the DC power
source (e.g. battery pack or vehicle) as described in the option's installation
guide.
2. Set the switch on the DC power supply to "I". A green LED indicates that the
DC power supply is ready for operation.
3. In case of overvoltage or undervoltage, the DC power supply switches off
automatically. This state is indicated by means of a red LED.
4. Set the power switch on the rear panel to "I".
5. Press the ON/OFF key on the front panel. A green LED above the ON/OFF
key indicates that the instrument is ready for operation.
Switching the DC power supply off
1. Press the ON/OFF key on the front panel of the R&S
FSVA/FSV and wait until
the instrument has shut down.
2. Switch off the power switch on the rear panel of the R&S
FSVA/FSV.