Annex
NRP2
391
User Manual 1173.9140.02 ─ 07
Integration interval
EXCLude:STARt
EXCLude:STOP
Power
Time
Fig. 1-8: Exclude parameter
Moreover, in Timeslot and Timegate it is possible to exclude time domains also during
measurement. (
...[:EXCL]:MID:OFFS
and
...:[:EXCL]:MID:TIME)
.
Example:
GSM measurement: The average power in the fifth GSM slot is to be measured. The
burst power fails by 50 µs and this range is to be excluded from the measurement..
Solution: A 50 µs-Fence is defined.. The sample values in this interval are not included
in the measurement:
*RST
SENS:FUNC "POW:TGAT:AVG"
SENS:TGAT:SEL 2
'select second gate
SENS:TGAT2:OFFS 2.328 MS
'GSM measurement in the fifth GSM slot
SENS:TGAT2:TIME 527 US
'length of a GSM burst
SENS:TGAT2:MID ON
'activate Fence
SENS:TGAT2:MID:OFFS 2.528 MS
'exclude interval starts, 200 µs after slot start
SENS:TGAT2:MID:TIME 50 US
'exclude interval starts, 200 µs after slot start
INIT
FETCh?