
Remote Control - Commands
NRP2
326
User Manual 1173.9140.02 ─ 07
SYSTem:SENSor<[1]..4>:INFO?
Returns information on a sensor. <string> is used to query a specific information item.
If called without parameters, the command returns all available information in the form
of a list of strings separated by commas. The permissible values for <string> depend
on the sensor. For details refer to the documentation of the corresponding sensor.
Parameters:
<string>
Range:
depending on sensor
Usage:
Query only
SYSTem:SENSor<[1]..4>:REBoot
Performs a power down reset of the selected sensor.
Usage:
Event
SYSTem:SENSor<[1]..4>:RESet
From the point of view of the R&S
NRP2 base unit, the sensors are stand-alone mea-
suring devices. They communicate with the R&S
NRP2 via a command set complying
with SCPI.
SYST:SENS<[1].48>:RES
prompts the base unit to send an
*RST
to the sensor in
question. Measurements in progress are interrupted.
Usage:
Event
SYSTem:SENSor<[1]..4>:TEST?
Performs a sensor test and returns a list of strings separated by commas. The contents
of this test protocol is sensor-specific. For its meaning, please refer to the sensor docu-
mentation.
Usage:
Query only
SYSTem:SPEed
<state>
The data processing speed of the R&S
NRP2 can be increased when
FAST
is
selected. The display is switched off and the measured values are no longer displayed
since the continuous update of the screen content requires computation time.
Parameters:
<state>
NORMal | FAST
*RST:
NORMal
SYSTem
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com