Getting Started
NRP2
33
User Manual 1173.9140.02 ─ 07
Fig. 2-9: Offset dialog box / Global activated
Tip:
Factors which are not, or only minimally, depending on frequency can be cor-
rected in this way.
7. Press the
↵
MENU hardkey twice to exit the dialog box and the menu.
If you have connected an attenuator and entered its attenuation, the R&S
NRP2
will display about the same value as before.
Fig. 2-10: Result window / Offset
In the note line of the result window, the
Δ
symbol indicates that global offset cor-
rection is active.
Performing relative power measurements
The R&S
NRP2 calculates and displays also the relative difference between a mea-
sured value and a reference value. The reference value can be a measured value that
has been saved or an arbitrary value that is entered.
1. Select the "Measurement" menu.
2. Open the "Measurement" menu with
↵
MENU.
Measuring average power (Cont Av mode)