Instrument Functions
NRP2
106
User Manual 1173.9140.02 ─ 07
Fig. 4-39: Function & Unit dialog box, sensors are operated in timeslot mode
Timeslot
Determines which timeslot is to be evaluated. Only one value for both channels can be
selected.
Eval.
This selection is only available in the "Trace & Statistics" mode. In the T'Slot Av mode,
evaluation is fixed to 'Avg' = average power in the timeslot.
Selects the measurand:
"Avg"
Average power in the timeslot
"Pk"
Peak power in the timeslot
"Pk/Avg"
Ratio of peak power to average power in the timeslot
Remote command:
Selection of display
DISPlay[:WINDow]<[1]..4>:AVALue
Reading out values
4.4.3.4
Setting the unit
Combining channels which are operated in different measurement modes
It is possible to define measurement functions based upon channels that are operated
in different measurement modes. This way, the average power within a specific time-
slot of a signal can be related to the average burst power of another signal. Many other
combinations are possible. The "Function & Unit" dialog box reflects this situation by
offering only those selection boxes that are needed to clearly define the measurands of
the combined channels.
Function & Unit dialog box / Units
The unit for indicating the results can be selected either directly in the "Measurement"
menu or in the "Function & Unit" dialog box.
Configuring measurements