Remote Control - Commands
NRP2
247
User Manual 1173.9140.02 ─ 07
Parameters:
<window_size>
NORMal | ZOOMed
NORMal
The window size is determined automatically and depends on
the total number of open windows.
ZOOMed
The selected window (
DISP:WIND:SEL[?]
) is set to maximum
size, regardless of the the number of visible windows. All other
open windows are covered.
*RST:
NORMal
DISPlay[:WINDow]<[1]..4>[:STATe]
<state>
Opens or closes a window. When a window is closed the other windows are automati-
cally expanded so that the free space is utilized. Likewise, open windows become
smaller when a new one is opened.
Parameters:
<state>
ON | OFF
*RST:
ON (Window 1) | OFF (Window 2,3,4)
DISPlay[:WINDow]<[1]..4>:TRACe:LOWer
<lower_limit>
Sets the lower limit for the power axis in the Trace mode. The value range depends on
the unit.
Table 6-13: The value range depends on the unit
Unit of measured value
W
DBM
DBUV
Value range
-999.999E3
..
999.999E3
-120
..
90
-13.010
..
136.990
Unit:
DBM | W | DBUV
Parameters:
<lower_limit>
float_value
*RST:
- 120 DB
Default unit: DBM
DISPlay[:WINDow]<[1]..4>:TRACe:UPPer
<upper_limit>
Sets the upper limit for the power axis in the Trace mode.
Value range:
The value range depends on the unit
.
Unit:
DBM | W | DBUV
DISPlay