Annex
NRP2
387
User Manual 1173.9140.02 ─ 07
The figure above shows the basic structure of the trigger system. It is possible to
improve the sequence of measurement using the commands from the
TRIGger
com-
mand system.
A.4.3 Logic relationship of command systems SENSe, CALCulate and
UNIT
The following section provides an overview of the measurement procedure of the
R&S
NRP2. The commands are mentioned as examples. These commands are descri-
bed in detail in
chapter 6, "Remote Control - Commands"
To perform measurements with the R&S
NRP2, the user should have a good under-
standing of the trigger system and be familiar with the result processing. The measured
values delivered by the sensors cannot be directly queried. As shown in figure below,
the raw data is first determined in the sensors taking into account the
SENSe
settings.
It is then transferred to the
CALCulate
blocks where the measured values of up to two
sensors are processed and made available to the user in the unit defined by
UNIT
.
CALC[1..8]:MATH:EXPR
- "SENS
n
"
- "SENS
n
- SENS
m
"
- "SENS
n
/ SENS
m
"
- "REFL"
- "SWR
- "RLOS"
CALC[1..8]:REL
UNIT[1..8]:POW
UNIT[1..8]:RAT
Data output:
output queue in
REMOTE mode
or display
in LOCAL mode
Measured values
("raw data")
CALCulate
UNIT
SENSe
SENS[1..4]:...
Fig. 1-6: Processing of measured values in the R&S
NRP2
The high-level commands and the
SENSe
block allow indices to be specified. These
indices refer to the connected sensors (indices 1 to 4) for the
SENSe
commands
whereas the indices of the high-level commands select the
CALCulate
blocks (indices
1 to 8). Depending on the configuration of the first
CALCulate
block,
FETCh?
(same
meaning as
FETCh1?
) may not deliver the measured value of sensor 1 in contrast to
the above examples. After
*RST
, the first
CALCulate
block delivers the measured
value of sensor 1 (
CALC1:MATH "(SENS1)"
) and the second one the measured
A.4.4 Measuring with low-level commands
Low-level commands allow the user to perform the most important modifications on the
measurement configuration. In the present context, low-level commands are all com-
Programming Measurements
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com